| Literature DB >> 24973310 |
Gregor Bruggmoser1, Rainer Saum2, Felicitas Saum2, Mark Gainey2, Christian Pychlau3, Ralf-Peter Kapsch4, Klemens Zink5.
Abstract
For the determination of absorbed dose to water D,using thermolumeniscence (TL) probes in a beam different from that used for calibration, correction factors for radiation type and radiation quality kE and kQ are needed. Values for kE and kQ for two different shapes of LiF probes (rods and disks) were obtained for high-energy photon and electron beams. The relation between the absorbed dose to the medium (water) D, measured by ion-chambers according to DIN 6800-2, 2008 and TL-probes having a (60)Co-calibration factor, leads for each shape and each batch of LiF probes to correction factors for radiation type and radiation quality kE and kQ.. The influence of the shape on the correction factor of the probes amounts in our experiment up to 2%. Therefore, it is recommended that the correction factors kE and kQ for rods and disks should be checked for each batch of LiF-detectors.Entities:
Keywords: Korrektionsfaktoren; LiF-Detektoren; LiF-detectors; Thermoluminescence-dosimetry; Thermolumineszenz-Dosimetrie; correction factors
Mesh:
Substances:
Year: 2014 PMID: 24973310 DOI: 10.1016/j.zemedi.2014.03.010
Source DB: PubMed Journal: Z Med Phys ISSN: 0939-3889 Impact factor: 4.820