Output calibration in solid water for high energy photon beams.

C S Reft1.   

Abstract

The AAPM Protocol recommends the use of water, polystyrene or acrylic media for measuring the output of high energy photon beams. It provides the appropriate restricted mass stopping powers and mass energy absorption coefficients for converting the dose to these media to dose to water. A water-equivalent solid has been developed for dosimetric applications [C. Constantinou, F. Attix, and B. Paliwal, Med. Phys. 9, 436 (1982)]. Calculated values for the restricted mass stopping powers and mass energy absorption coefficients have been published for this material [A. Ho and B. Paliwal, Med. Phys. 13, 403 (1986)]. The accuracy of these calculations was investigated by making output measurements, following the Protocol, with a Farmer type chamber in four materials for Co-60, 4, 6, 10, 18, and 24 MV photon beams. The results show that the scaled dose to water for the different media agree to better than 1%, and the analysis supports the methodology of the Protocol for obtaining the dose to water from the different media.

Entities:  

Mesh:

Substances:

Year:  1989        PMID: 2497318     DOI: 10.1118/1.596423

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Dosimetric impact of density variations in Solid Water 457 water-equivalent slabs.

Authors:  Dale W Litzenberg; Hanan Amro; Joann I Prisciandaro; Eduardo Acosta; Ian Gallagher; Don A Roberts
Journal:  J Appl Clin Med Phys       Date:  2011-04-22       Impact factor: 2.102

2.  Dosimetric evaluation of Plastic Water Diagnostic-Therapy.

Authors:  Ramani Ramaseshan; Kirpal Kohli; Fred Cao; Robert K Heaton
Journal:  J Appl Clin Med Phys       Date:  2008-04-29       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.