Patient radiation exposure during carotid artery stenting.

Michał-Goran Stanišić1, Natalia Majewska2, Marcin Makałowski2, Robert Juszkat2, Magdalena Błaszak3, Wacław Majewski4.   

Abstract

OBJECTIVES: The main purpose of this study was to document the radiation doses to patients during carotid stenting. MATERIAL AND
METHOD: Fluoroscopy and exposure time, air kerma and dose-area product during carotid artery stenting in 160 patients were retrospectively reviewed with regard to body mass index, degree of stenosis and use of cerebral protection devices.
RESULTS: Total air kerma was lower than 0.5 Gy in 80%, 0.5-1 Gy in 17% and higher than 1 Gy (maximum 1.2) in 3% of patients. Mean total dose-area product value for carotid stenting was 54 Gy cm(2). The mean air kerma (fluoroscopy), air kerma (exposure), total air kerma and dose-area product (fluoroscopy), dose-area product (exposure), total dose-area product of patients with body mass index within the range 25-29.9 and with body mass index >30 were significantly increased compared to that of patients with body mass index 18-24.9 (H = 40.2, df = 2; p = 0.0000001 and p = 0.000003, respectively).
CONCLUSION: Carotid artery stenting is a relatively safe radiological procedure in terms of the radiation dose acquired by the patient. The main factors contributing to possible radiation overdosing are body mass index value and complexity of the carotid lesion. Proper preoperative planning in obese and complicated patients may reduce the fluoroscopy time and contribute to reduced dose acquisition.
© The Author(s) 2014 Reprints and permissions: sagepub.co.uk/journalsPermissions.nav.

Entities:  

Keywords:  Carotid artery stenting; air kerma; dose-area product; radiation exposure; radiological safety

Mesh:

Year:  2014        PMID: 24966273     DOI: 10.1177/1708538114540641

Source DB:  PubMed          Journal:  Vascular        ISSN: 1708-5381            Impact factor:   1.285


  1 in total

1.  Comparison between anteroposterior and oblique "Scotty dog" approach during S1 transforaminal epidural steroid injection: A randomized controlled trial.

Authors:  Ryung A Kang; Woo Seog Sim; Ji Won Choi; Sehee Kang; Seungwon Lee; Hyeon Sook Jee; Eun Jung Oh; Yunghun Kim; Justin Sangwook Ko
Journal:  Medicine (Baltimore)       Date:  2020-10-23       Impact factor: 1.817

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.