BACKGROUND AND
PURPOSE: Proper cleaving of reusable laser fibers is needed to maintain optimal functionality. This study quantifies the effect of different cleaving tools on power output of the holmium laser fiber and demonstrates morphologic changes using microscopy.
MATERIALS AND METHODS: The uncleaved tips of new 272 μm reusable laser fibers were used to obtain baseline power transmission values at 3 W (0.6 J, 5 Hz). Power output for each of four cleaving techniques-11-blade scalpel, scribe pen cleaving tool, diamond cleaving wheel, and suture scissors-was measured in a single-blinded fashion. Dispersion of light from the fibers was compared with manufacturer specifications and rated as "ideal," "acceptable," or "unacceptable" by blinded reviewers. The fiber tips were also imaged using confocal and scanning electron microscopy. Independent samples Kruskal-Wallis test and chi square were used for statistical analysis (α<0.05).
RESULTS: New uncleaved fiber tips transmitted 3.04 W of power and were used as a reference (100%). The scribe pen cleaving tool produced the next highest output (97.1%), followed by the scalpel (83.4%), diamond cleaving wheel (77.1%), and suture scissors (61.7%), a trend that was highly significant (P<0.001). On pairwise comparison, no difference in power output was seen between the uncleaved fiber tips and those cleaved with the scribe pen (P=1.0). The rating of the light dispersion patterns from the different cleaving methods followed the same trend as the power output results (P<0.001). Microscopy showed that the scribe pen produced small defects along the fiber cladding but maintained a smooth, flat core surface. The other cleaving techniques produced defects on both the core and cladding.
CONCLUSION: Cleaving techniques produce a significant effect on the initial power transmitted by reusable laser fibers. The scribe pen cleaving tool produced the most consistent and highest average power output.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 24964275     DOI: 10.1089/end.2014.0175

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  4 in total

1.  Impact of laser fiber tip cleavage on power output for ureteroscopy and stone treatment.

Authors:  M Haddad; E Emiliani; Y Rouchausse; F Coste; L Berthe; S Doizi; S Buttice; B Somani; O Traxer
Journal:  World J Urol       Date:  2017-05-30       Impact factor: 4.226

2.  The effect of laser fiber on the damage of the working channel of a flexible ureteroscope.

Authors:  Kantima Jongjitaree; Chaiyan Subpayakorn; Tawatchai Taweemonkongsap; Sunai Leewansangtong; Sittiporn Srinualnad; Ekkarin Chotikawanich
Journal:  Heliyon       Date:  2020-11-27

Review 3.  Holmium Laser Lithotripsy in the New Stone Age: Dust or Bust?

Authors:  Ali H Aldoukhi; William W Roberts; Timothy L Hall; Khurshid R Ghani
Journal:  Front Surg       Date:  2017-09-29

Review 4.  Advances in Lasers for the Treatment of Stones-a Systematic Review.

Authors:  Peter Kronenberg; Bhaskar Somani
Journal:  Curr Urol Rep       Date:  2018-05-17       Impact factor: 3.092

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.