| Literature DB >> 24954069 |
N B Sing1, A Mostavan, E Hamzah, D Mantovani, H Hermawan.
Abstract
This article reports a degradation study that was done on stent prototypes made of biodegradable Fe35Mn alloy in a simulated human coronary arterial condition. The stent degradation was observed for a short-term period from 0.5 to 168 h, which simulates the early period of stenting procedure. Potentiodynamic polarization and electrochemical impedance spectroscopy were used to quantify degradation rate and surface property of the stents. Results showed that signs of degradation were visible on both crimped and expanded stents after 1 h of test, mostly located on the stent's curvatures. The degradation rate of stent was higher compared to that of the original alloy, indicating the surface altering effect of stent fabrication processing to degradation. A single oxide layer was formed and detected as a porous structure with capacitive behavior. Expanded stents exhibited lower polarization resistance compared to the nonexpanded ones, indicating the cold work effect of expansion procedure to degradation.Entities:
Keywords: biodegradable metals; biodegradation; electrochemical impedance spectroscopy; stents
Mesh:
Substances:
Year: 2014 PMID: 24954069 DOI: 10.1002/jbm.b.33242
Source DB: PubMed Journal: J Biomed Mater Res B Appl Biomater ISSN: 1552-4973 Impact factor: 3.368