Literature DB >> 24947967

Monte Carlo photon beam modeling and commissioning for radiotherapy dose calculation algorithm.

A Toutaoui1, S Ait chikh2, N Khelassi-Toutaoui3, B Hattali4.   

Abstract

The aim of the present work was a Monte Carlo verification of the Multi-grid superposition (MGS) dose calculation algorithm implemented in the CMS XiO (Elekta) treatment planning system and used to calculate the dose distribution produced by photon beams generated by the linear accelerator (linac) Siemens Primus. The BEAMnrc/DOSXYZnrc (EGSnrc package) Monte Carlo model of the linac head was used as a benchmark. In the first part of the work, the BEAMnrc was used for the commissioning of a 6 MV photon beam and to optimize the linac description to fit the experimental data. In the second part, the MGS dose distributions were compared with DOSXYZnrc using relative dose error comparison and γ-index analysis (2%/2 mm, 3%/3 mm), in different dosimetric test cases. Results show good agreement between simulated and calculated dose in homogeneous media for square and rectangular symmetric fields. The γ-index analysis confirmed that for most cases the MGS model and EGSnrc doses are within 3% or 3 mm.
Copyright © 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  EGSnrc; Monte Carlo; TPS commissioning

Mesh:

Substances:

Year:  2014        PMID: 24947967     DOI: 10.1016/j.ejmp.2014.05.007

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  5 in total

1.  Monte Carlo simulation using PRIMO code as a tool for checking the credibility of commissioning and quality assurance of 6 MV TrueBeam STx varian LINAC.

Authors:  M Arif Efendi; Amporn Funsian; Thawat Chittrakarn; Tripob Bhongsuwan
Journal:  Rep Pract Oncol Radiother       Date:  2019-12-20

2.  Monte carlo model and output factors of elekta infinity™ 6 and 10 MV photon beam.

Authors:  Sitti Yani; Indra Budiansah; Mohamad Fahdillah Rhani; Freddy Haryanto
Journal:  Rep Pract Oncol Radiother       Date:  2020-04-28

3.  Monte-Carlo simulation of the Siemens Artiste linear accelerator flat 6 MV and flattening-filter-free 7 MV beam line.

Authors:  Alemeh Sadrollahi; Frank Nuesken; Norbert Licht; Christian Rübe; Yvonne Dzierma
Journal:  PLoS One       Date:  2019-01-08       Impact factor: 3.240

4.  The commissioning and validation of Monaco treatment planning system on an Elekta VersaHD linear accelerator.

Authors:  Jeffrey E Snyder; Daniel E Hyer; Ryan T Flynn; Amanda Boczkowski; Dongxu Wang
Journal:  J Appl Clin Med Phys       Date:  2018-12-07       Impact factor: 2.102

5.  Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom.

Authors:  Yuxi Pan; Ruijie Yang; Jun Li; Xile Zhang; Lu Liu; Junjie Wang
Journal:  J Appl Clin Med Phys       Date:  2018-03-30       Impact factor: 2.102

  5 in total