Literature DB >> 24941931

Evaluation of the spectra Optia® mononuclear cell collection procedure in multiple myeloma patients.

Gwynn Long1, Edmund K Waller, Steven Gregurek, Guido Tricot, Susanne Marschner, Jerry Bill.   

Abstract

BACKGROUND: Peripheral blood stem cell (PBSC) rescue following myeloablative therapy is a mainstay of cancer therapy. To evaluate the ability of the Spectra Optia Apheresis System (SO), a newly developed apheresis device, the device was studied in multiple myeloma patients undergoing a first autologous PBSC transplant. AIM: To demonstrate that neutrophil recovery was not inferior to historical controls when SO harvested PBSCs were reinfused following myeloablative therapy.
METHODS: Multiple myeloma patients were mobilized according to the standard practice at four clinical sites. Following mobilization, MNC collections were performed on the SO. The collected cells were cryopreserved and reinfused following myeloablative chemotherapy. Neutrophil recovery defined by an absolute neutrophil count exceeding 500/μL (ANC500) was compared to historical data for patients transplanted following apheresis using the COBE Spectra (CS) device.
RESULTS: The median day to neutrophil recovery was 12 days (range 10-14 days), with no significant difference in engraftment comparing patients transplanted with stem cells collected using the SO versus historical cohort of patients collected with the CS. CD34+ cell and MNC collection efficiency (CE) were 69.3% and 65.0% for the SO and CS, respectively. Platelet CE, product hematocrit and product granulocytes (as % of WBCs) using the SO were 21%, 2.3% and 28%, respectively. There were no device-related severe adverse events.
CONCLUSIONS: The study's results confirm that the Spectra Optia Apheresis System's MNC Collection Protocol is safe and effective for its intended use and that engraftment kinetics of cells collected by SO is not inferior to the CS System.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  apheresis; autologous transplantation; mononuclear cell collection; neutrophil recovery; peripheral blood stem-cell transplant

Mesh:

Substances:

Year:  2014        PMID: 24941931     DOI: 10.1002/jca.21341

Source DB:  PubMed          Journal:  J Clin Apher        ISSN: 0733-2459            Impact factor:   2.821


  2 in total

1.  Collection of hematopoietic CD34 stem cells in rhesus macaques using Spectra Optia.

Authors:  Lynn D Haynes; Jennifer Coonen; Jennifer Post; Kevin Brunner; Debra Bloom; Peiman Hematti; Dixon B Kaufman
Journal:  J Clin Apher       Date:  2016-08-31       Impact factor: 2.821

2.  Comparison of Two Apheresis Systems of COBE and Optia for Autologous Peripheral Blood Stem Cell Collection.

Authors:  Se Na Lee; Ji Yeon Sohn; Jung Hee Kong; Hyeon Seok Eom; Hyewon Lee; Sun Young Kong
Journal:  Ann Lab Med       Date:  2017-07       Impact factor: 3.464

  2 in total