Individually adapted tube current selection and contrast medium injection protocol of coronary CT angiography based on test bolus parameters: a feasibility study.

Xiaomei Zhu1, Zhaofei Shi1, Yinsu Zhu1, Wangyan Liu1, Guanyu Yang2, Tongfu Yu1, Lijun Tang3, Yi Xu1.   

Abstract

BACKGROUND: Test bolus is mostly used to determine the starting point of a full cardiac scanning with respect to injection of a larger bolus of contrast material. So far there are limited data demonstrating the feasibility of using information obtained from a test bolus to adjust contrast delivery protocols and tube current individually during coronary computed tomography angiography (CCTA).
PURPOSE: To evaluate the feasibility of individually adapted tube current selection and contrast injection protocols of CCTA based on test bolus parameters.
MATERIAL AND METHODS: Test bolus followed by CCTA was performed in 93 patients at 100 kV and in 81 patients at 120 kV, respectively. Simulated attenuation of the descending aorta (SimDA) of CCTA was calculated at a fixed contrast injection rate of 4 mL/s. Univariate and multivariate comparisons were performed to identify associations of SimDA and image noise of CCTA (NoiseCCTA) with test bolus information and patient-related factors including body weight (BW), body mass index (BMI), and body surface area (BSA).
RESULTS: Compared with BW, BMI, and BSA, SimDA was more closely related to the peak time of left ventricle and peak enhancement of right ventricle obtained from test bolus (r = 0.495 and r = 0.642 for 100 and 120 kV protocol, respectively). Similarly, NoiseTB was much more closely related to NoiseCCTA (r = 0.740 and r = 0.630 for 100 and 120 kV protocol, respectively) when compared with BW, BMI, and BSA.
CONCLUSION: It is feasible to individually adapt tube current and contrast injection protocol of CCTA based on the information of test bolus. © The Foundation Acta Radiologica 2014 Reprints and permissions: sagepub.co.uk/journalsPermissions.nav.

Entities:  

Keywords:  Coronary CT angiography; contrast enhancement; descending aorta; image noise; test bolus

Mesh:

Substances:

Year:  2014        PMID: 24938658     DOI: 10.1177/0284185114536156

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  2 in total

1.  Coronary angiography using second-generation dual source computed tomography: Feasibility of low dose and low flow rate to achieve appropriate individual contrast enhancement using a test bolus-based contrast medium protocol-A CONSORT compliant article.

Authors:  Xiaomei Zhu; Yusheng Yu; Dinghu Xu; Hong Zhang; Lijun Tang
Journal:  Medicine (Baltimore)       Date:  2018-07       Impact factor: 1.889

2.  Automatic tube potential selection with tube current modulation in coronary CT angiography: Can it achieve consistent image quality among various individuals?

Authors:  Xiao-Ping Wang; Xiao-Mei Zhu; Yin-Su Zhu; Wang-Yan Liu; Xiao-Han Yang; Wei-Wei Huang; Yi Xu; Li-Jun Tang
Journal:  Exp Ther Med       Date:  2018-05-11       Impact factor: 2.447

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.