Ji Hyun Youk1, Eun Ju Son1, Ah Young Park1, Jeong-Ah Kim1.
Abstract
PURPOSE: To evaluate and compare the performance of shear-wave elastography (SWE) for breast masses using the local shear wave speed (m/sec) vs. Young modulus (kPa).Entities:
Keywords: Breast; Sonoelastography; Ultrasonography, mammary
Year: 2013 PMID: 24936493 PMCID: PMC4058968 DOI: 10.14366/usg.13005
Source DB: PubMed Journal: Ultrasonography ISSN: 2288-5919
Fig. 1.B-mode and shear-wave elastography (SWE) images in a 46-year-old woman with a pathologically proven invasive ductal carcinoma.
A. SWE (right) and B-mode images (left) on split-screen mode show a 20-mm, irregular, spiculated mass with red, heterogeneous elasticity. B. The mean, maximum, and standard deviation of elasticity values were measured in kPa (left) and in m/sec (right) by placing the region-ofinterest (ROI) over the stiffest part of the lesion. C. The ROI for measuring the elasticity value (wSD) was placed to include the whole breast lesion and the stiffest part of the lesion.
Elasticity values in benign and malignant breast lesions
| Variable | kPa | m/sec | ||||||
|---|---|---|---|---|---|---|---|---|
| Emax | Emean | SD | wSD | Emax | Emean | SD | wSD | |
| Benign | 40.6±19.5 | 34.8±17.7 | 3.6±2.0 | 6.7±3.6 | 3.6±0.9 | 3.3±0.8 | 1.1±0.3 | 1.5±0.4 |
| Malignant | 155.1±67.2 | 140.7±58.5 | 14.1±9.6 | 30.8±16.1 | 7.1±1.5 | 6.7±1.5 | 2.1±0.7 | 3.1±0.9 |
| P-value | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 |
Values are presented as mean±SD.
Emax, maximum elasticity; Emean, mean elasticity value; wSD, SD of elasticity value of the whole breast lesion.
Diagnostic performance of elasticity values for breast cancer
| Variable | Sensitivity (%) | Specificity (%) | AUC (95% CI) | |
|---|---|---|---|---|
| Emax | kPa | 91.8 | 88.9 | 0.960 (0.911 to 0.987) |
| m/s | 93.9 | 88.9 | 0.976 (0.933 to 0.995) | |
| P-value | >0.990 | >0.990 | 0.345 | |
| Emean | kPa | 89.8 | 93.8 | 0.974 (0.930 to 0.994) |
| m/s | 83.7 | 98.8 | 0.974 (0.929 to 0.994) | |
| P-value | 0.250 | 0.125 | 0.834 | |
| SD | kPa | 79.6 | 95.1 | 0.916 (0.854 to 0.957) |
| m/s | 85.7 | 87.7 | 0.916 (0.854 to 0.957) | |
| P-value | 0.250 | 0.031 | 0.932 | |
| wSD | kPa | 89.4 | 92.5 | 0.964 (0.915 to 0.989) |
| m/s | 89.4 | 91.3 | 0.960 (0.909 to 0.987) | |
| P-value | >0.990 | >0.990 | 0.036 |
AUC, area under the receiver operating characteristic curve; CI, confidence interval; Emax, maximum elasticity, SD, standard deviation; Emean, mean elasticity value; wSD, SD of elasticity value of the whole breast lesion.
Fig. 2.Receiver operating characteristic curves for the standard deviation of the elasticity value of the whole breast lesion (wSD) in kPa and m/sec.
The areas under the curves were significantly different using kPa and m/sec (0.964 vs. 0.960, P=0.036).