S Lenschow1, B Schliemann, M Schulze, M Raschke, C Kösters. 1. Department of Trauma, Hand and, Reconstructive Surgery, Wilhelms University Muenster, Waldeyerstraße 1, 48149, Münster, Germany, simon.lenschow@ukmuenster.de.
Abstract
INTRODUCTION: The aim of this study was to compare the biomechanical properties of tibial fixation of a free tendon graft in ACL reconstruction using the Shim, a new wedge-shaped implant, in an outside-in technique to fixation by the Shim used in an inside-out technique and fixation by interference screw in a porcine model. MATERIALS AND METHODS: Porcine tibia and flexor tendons were used. In Group 1, the Shim was applied outside-in. In Group 2, the Shim was inserted inside-out. In the Group 3, an 8-mm interference screw was used. Ten specimens were tested in each group. Load-to-failure, elongation, stiffness and failure mode were recorded. Cyclic loading was performed between 5 and 250 N for 1,000 cycles, followed by a load to failure testing. RESULTS: Mean maximum load-to-failure was 629.53 N in Group 1,648.54 N in Group 2 and 749.53 N in Group 3. There was no significant difference between the groups. Stiffness varied between 127.34 N/mm in Group 1, 151.27 N/mm in Group 2 and 182.25 N/mm in Group 3. No significant differences were found between outside-in Shim and interference screw fixation. No significant difference was found for elongation among the three groups. The main failure mode was a rupture of the tendon in the IFS group and a slippage of either the implant or the tendon in both groups using the Shim. CONCLUSIONS: As no statistically significant difference could be seen concerning load to failure, stiffness and elongation between the inside-out and the outside-in techniques, the Shim can be used for tibial fixation in an outside-in or inside-out technique depending on the preference of the surgeon. To prevent slippage of the graft a hybrid fixation should be considered.
INTRODUCTION: The aim of this study was to compare the biomechanical properties of tibial fixation of a free tendon graft in ACL reconstruction using the Shim, a new wedge-shaped implant, in an outside-in technique to fixation by the Shim used in an inside-out technique and fixation by interference screw in a porcine model. MATERIALS AND METHODS: Porcine tibia and flexor tendons were used. In Group 1, the Shim was applied outside-in. In Group 2, the Shim was inserted inside-out. In the Group 3, an 8-mm interference screw was used. Ten specimens were tested in each group. Load-to-failure, elongation, stiffness and failure mode were recorded. Cyclic loading was performed between 5 and 250 N for 1,000 cycles, followed by a load to failure testing. RESULTS: Mean maximum load-to-failure was 629.53 N in Group 1,648.54 N in Group 2 and 749.53 N in Group 3. There was no significant difference between the groups. Stiffness varied between 127.34 N/mm in Group 1, 151.27 N/mm in Group 2 and 182.25 N/mm in Group 3. No significant differences were found between outside-in Shim and interference screw fixation. No significant difference was found for elongation among the three groups. The main failure mode was a rupture of the tendon in the IFS group and a slippage of either the implant or the tendon in both groups using the Shim. CONCLUSIONS: As no statistically significant difference could be seen concerning load to failure, stiffness and elongation between the inside-out and the outside-in techniques, the Shim can be used for tibial fixation in an outside-in or inside-out technique depending on the preference of the surgeon. To prevent slippage of the graft a hybrid fixation should be considered.
Authors: Benedikt Schliemann; Niklas Grüneweller; Daiwei Yao; Clemens Kösters; Simon Lenschow; Steffen B Roßlenbroich; Michael J Raschke; Andre Weimann Journal: Int Orthop Date: 2015-10-03 Impact factor: 3.075
Authors: J Glasbrenner; C Domnick; M J Raschke; T Willinghöfer; C Kittl; P Michel; D Wähnert; Mirco Herbort Journal: Knee Surg Sports Traumatol Arthrosc Date: 2018-10-27 Impact factor: 4.342