Comparison of outside-in and inside-out technique for tibial fixation of a soft-tissue graft in ACL reconstruction using the Shim technique.

S Lenschow1, B Schliemann, M Schulze, M Raschke, C Kösters.   

Abstract

INTRODUCTION: The aim of this study was to compare the biomechanical properties of tibial fixation of a free tendon graft in ACL reconstruction using the Shim, a new wedge-shaped implant, in an outside-in technique to fixation by the Shim used in an inside-out technique and fixation by interference screw in a porcine model.
MATERIALS AND METHODS: Porcine tibia and flexor tendons were used. In Group 1, the Shim was applied outside-in. In Group 2, the Shim was inserted inside-out. In the Group 3, an 8-mm interference screw was used. Ten specimens were tested in each group. Load-to-failure, elongation, stiffness and failure mode were recorded. Cyclic loading was performed between 5 and 250 N for 1,000 cycles, followed by a load to failure testing.
RESULTS: Mean maximum load-to-failure was 629.53 N in Group 1,648.54 N in Group 2 and 749.53 N in Group 3. There was no significant difference between the groups. Stiffness varied between 127.34 N/mm in Group 1, 151.27 N/mm in Group 2 and 182.25 N/mm in Group 3. No significant differences were found between outside-in Shim and interference screw fixation. No significant difference was found for elongation among the three groups. The main failure mode was a rupture of the tendon in the IFS group and a slippage of either the implant or the tendon in both groups using the Shim.
CONCLUSIONS: As no statistically significant difference could be seen concerning load to failure, stiffness and elongation between the inside-out and the outside-in techniques, the Shim can be used for tibial fixation in an outside-in or inside-out technique depending on the preference of the surgeon. To prevent slippage of the graft a hybrid fixation should be considered.

Entities:  

Mesh:

Year:  2014        PMID: 24935662     DOI: 10.1007/s00402-014-2029-1

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  5 in total

1.  Biomechanical evaluation of different surgical techniques for treating patellar tendon ruptures.

Authors:  Benedikt Schliemann; Niklas Grüneweller; Daiwei Yao; Clemens Kösters; Simon Lenschow; Steffen B Roßlenbroich; Michael J Raschke; Andre Weimann
Journal:  Int Orthop       Date:  2015-10-03       Impact factor: 3.075

2.  Adjustable buttons for ACL graft cortical fixation partially fail with cyclic loading and unloading.

Authors:  J Glasbrenner; C Domnick; M J Raschke; T Willinghöfer; C Kittl; P Michel; D Wähnert; Mirco Herbort
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2018-10-27       Impact factor: 4.342

3.  Enhanced Bone-Tendon-Bone Approach for Open Anterior Cruciate Ligament Replacement With Conservation of the Joint Capsule.

Authors:  Sebastian Gottfried Walter; Tom Sascha Thomas; Luca Tafuro; Wolfram Thomas
Journal:  Arthrosc Tech       Date:  2015-10-26

4.  Categorize the existing clamps used for tensile test of human graft- a systematic review.

Authors:  Denes Farago; Blanka Kozma; Rita Maria Kiss
Journal:  BMC Musculoskelet Disord       Date:  2022-07-25       Impact factor: 2.562

5.  Comparison of the mechanical properties and mechanical damages to tendon tissue in three suspensory fixation techniques.

Authors:  Toshiya Kano; Masakazu Ishikawa; Yuki Ogawa; Hiroyuki Akebono; Atsushi Sugeta; Masato Ohmi; Goki Kamei; Atsuo Nakamae; Nobuo Adachi
Journal:  Asia Pac J Sports Med Arthrosc Rehabil Technol       Date:  2022-07-31
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.