Literature DB >> 24930649

The accuracy and uncertainty of a sheet-type portable monitor as a screening device to identify obstructive sleep apnea-hypopnea syndrome.

Masanori Tsukahara1, Seiichiro Sakao, Takayuki Jujo, Takayuki Sakurai, Jiro Terada, Reiko Kunii, Nobuhiro Tanabe, Koichiro Tatsumi.   

Abstract

OBJECTIVE: Laboratory-based polysomnography (PSG) is the gold standard for diagnosing obstructive sleep apnea-hypopnea syndrome (OSAHS), but it is expensive and requires overnight hospitalization. Recently, a sheet-shaped breath detection monitor, the SD-101, has been developed, and several reports have so far demonstrated the screening accuracy of this device. The aim of this study was to assess the accuracy and the uncertainty of this device.
METHODS: A total of 101 suspected OSAHS patients underwent simultaneous examinations with PSG and the SD-101.
RESULTS: There was a statistically significant relationship between the respiratory disturbance index (RDI) by the SD-101 and the apnea-hypopnea index (AHI) by PSG. At an RDI cutoff of 14 episodes per hour, the sensitivity and specificity to detect an AHI ≥20 episodes per hour were 90.2% and 90.0%, respectively. To reduce the influence of sleep efficiency, the time in bed (TIB) obtained from PSG, instead of the total seep time (TST), was used to calculate the AHI from the PSG data. There was also a statistically significant relationship between the RDI and AHI for the TIB. Moreover, it was suggested that arousal index and TIB were likely associated with false-negative and/or false-positive results.
CONCLUSION: Although the present study demonstrated a close relationship between the RDI and the AHI, use of the SD-101 to examine symptomatic OSAHS patients should be performed with a full understanding of its incapability to detect the sleep state, including arousal reaction and the existence of false respiratory events caused by body movements.

Entities:  

Mesh:

Year:  2014        PMID: 24930649     DOI: 10.2169/internalmedicine.53.2208

Source DB:  PubMed          Journal:  Intern Med        ISSN: 0918-2918            Impact factor:   1.271


  3 in total

1.  Diagnostic accuracy of level IV portable sleep monitors versus polysomnography for obstructive sleep apnea: a systematic review and meta-analysis.

Authors:  Lusine Abrahamyan; Yeva Sahakyan; Suzanne Chung; Petros Pechlivanoglou; Joanna Bielecki; Steven M Carcone; Valeria E Rac; Michael Fitzpatrick; Murray Krahn
Journal:  Sleep Breath       Date:  2018-01-09       Impact factor: 2.816

2.  Digital Health and Sleep-Disordered Breathing: A Systematic Review and Meta-Analysis.

Authors:  Talita Rosa; Kersti Bellardi; Alonço Viana; Yifei Ma; Robson Capasso
Journal:  J Clin Sleep Med       Date:  2018-09-15       Impact factor: 4.062

3.  Validation of a novel sleep-monitoring system for diagnosing obstructive sleep apnea: A comparison with polysomnography.

Authors:  Lili Meng; Huajun Xu; Jian Guan; Hongliang Yi; Hongmin Wu; Shankai Yin
Journal:  Exp Ther Med       Date:  2016-09-20       Impact factor: 2.447

  3 in total