| Literature DB >> 24930405 |
C Vogne1, G Prod'hom, K Jaton, L A Decosterd, G Greub.
Abstract
Carbapenemases should be accurately and rapidly detected, given their possible epidemiological spread and their impact on treatment options. Here, we developed a simple, easy and rapid matrix-assisted laser desorption ionization-time of flight (MALDI-TOF)-based assay to detect carbapenemases and compared this innovative test with four other diagnostic approaches on 47 clinical isolates. Tandem mass spectrometry (MS-MS) was also used to determine accurately the amount of antibiotic present in the supernatant after 1 h of incubation and both MALDI-TOF and MS-MS approaches exhibited a 100% sensitivity and a 100% specificity. By comparison, molecular genetic techniques (Check-MDR Carba PCR and Check-MDR CT103 microarray) showed a 90.5% sensitivity and a 100% specificity, as two strains of Aeromonas were not detected because their chromosomal carbapenemase is not targeted by probes used in both kits. Altogether, this innovative MALDI-TOF-based approach that uses a stable 10-μg disk of ertapenem was highly efficient in detecting carbapenemase, with a sensitivity higher than that of PCR and microarray.Entities:
Keywords: Carbapenem resistance; Gram-negative bacilli; MALDI-TOF MS; MS-MS spectrometry; PCR; microarray
Mesh:
Substances:
Year: 2014 PMID: 24930405 DOI: 10.1111/1469-0691.12715
Source DB: PubMed Journal: Clin Microbiol Infect ISSN: 1198-743X Impact factor: 8.067