| Literature DB >> 24914673 |
José Fábio Guastelli Gastaldo1, Angélica Castro Pimentel2, Maria Helena Gomes1, Wilson Roberto Sendyk2, Dalva Cruz Laganá1.
Abstract
The aim of study was to assess the stress around 10/13/15-mm implants in the mandibular area with a 15-mm cantilevered acrylic-resin-coated prostheses following the application force, using the photoelasticity method. Three photoelastic mandibular models were created containing 10-, 13-, and 15-mm implants in length and 3.75 mm in diameter. The implants had bore internal hex connections and were placed parallel to the intermental region. Abutments with 1-mm high cuffs were placed over the implants, and a single cobalt/chrome metallic prosthesis with a 15-mm cantilever, coated with thermoplastic acrylic resin, was placed on top. Loads of 1.0 and 3.0 bars were applied, and the images were photographed and assessed by photoelasticity method. The greatest stress levels were observed for the 10-mm implants. The stress pattern was the same regardless of implant length; only the magnitude of the stress along the implant body revealed changes. Increased implant length played a role in reducing stress on the investigated area of the model, and the 15-mm implants exhibited the best performance in regard to stress distribution. The highest stress levels were found in the implants closest to the cantilever and the central implant. The longest implants were more favorable in regard to the stress distribution on the peri-implant support structures in the 15-mm cantilevered prosthesis under loads.Entities:
Keywords: dental implants; load; mandible; stress
Mesh:
Substances:
Year: 2014 PMID: 24914673 DOI: 10.1563/AAID-JOI-D-13-00139
Source DB: PubMed Journal: J Oral Implantol ISSN: 0160-6972 Impact factor: 1.779