| Literature DB >> 24904244 |
Craig L Bull1, Michael W Johnson1, Hayrullo Hamidov2, Kazuki Komatsu2, Malcolm Guthrie2, Matthias J Gutmann3, John S Loveday2, Richard J Nelmes2.
Abstract
A robust and comprehensive method for determining the orientation matrix of a single-crystal sample using the neutron Laue time-of-flight (TOF) technique is described. The new method enables the measurement of the unit-cell parameters with an uncertainty in the range 0.015-0.06%, depending upon the crystal symmetry and the number of reflections measured. The improved technique also facilitates the location and integration of weak reflections, which are often more difficult to discern amongst the increased background at higher energies. The technique uses a mathematical model of the relative positions of all the detector pixels of the instrument, together with a methodology that establishes a reproducible reference frame and a method for determining the parameters of the instrument detector model. Since all neutron TOF instruments require precise detector calibration for their effective use, it is possible that the method described here may be of use on other instruments where the detector calibration cannot be determined by other means.Entities:
Keywords: calibration; neutron diffraction; neutron instruments; time-of-flight
Year: 2014 PMID: 24904244 PMCID: PMC4038798 DOI: 10.1107/S1600576714006657
Source DB: PubMed Journal: J Appl Crystallogr ISSN: 0021-8898 Impact factor: 3.304
Figure 1A schematic diagram of the SXD instrument. The neutrons leave the target and are energy-shifted by the water moderator. The size of the beam impinging on the sample position is defined by the adjustable BN jaws. The sample position is surrounded by 11 detectors, six in the equatorial plane and five below, maximizing the volume of reciprocal space that can be measured with one sample orientation. The frame that supports the detectors is shown in more detail in Fig. 2 ▶. Using the ISIS in-house software (SXD2001), the centre of each detector is defined by angles from the incident beam direction, denoted γ in the horizontal plane and νM in the vertical plane.
Figure 2(Not to common scale) (Top) A schematic diagram of the frame of reference defining the z direction. (Bottom) A schematic diagram of the frame of reference defining the x and y directions. A is the frame of SXD upon which are mounted the detectors and the CCR. Upon the SXD frame are two dowel pins labelled B onto which are located the tank of the CCR. From a datum point which is the surface labelled C, a position 300 mm below this is defined as the sample position D. A perpendicular line from surface C defines the z direction of the frame of reference. E is the suggested incident beam direction of SXD, F are the jaws defining the SXD beam size, G are kinematic mounts defining the SXD frame relative to the moderator and target and hence E, H is the rotation mechanism which rotates the crystal around the origin of the axis, and I are pins used to define the x direction, which is defined to pass through a line that can be drawn through these two points. The y direction is defined as being perpendicular to a plane which is parallel to that defined by the datum surface labelled C. J is additional beam-defining collimation.
Figure 3Detector 5, showing the pixel positions within the detector coordinate system (red) and the instrument coordinate system (blue).
A list of all the SXD model parameters, together with their uncertainties, nominal engineering values, origin and units
The origins of the parameter values were (a) metrology, (b) powder measurements, (c) ‘equatorial line’ determination and (d) refinement of known KDP single-crystal reflection data (see §6).
| Parameter | Model value | Nominal value | Origin/units | Parameter | Model value | Nominal value | Origin/units |
|---|---|---|---|---|---|---|---|
| px, pz | 3.000 (1) | 3.0 |
| αn | 0.00043 (2) | 0.0 |
|
|
| 8297.3 (6) | 8300 |
| βn | −0.0003 (4) | 0.0 |
|
|
| −176.1 (2) | −178.5 |
|
| 177.29 (20) | 178.5 |
|
|
| 139.43 (14) | 136.98 |
|
| −138.79 (19) | −136.98 |
|
|
| −0.73 (9) | 0.0 |
|
| −2.04 (8) | 0.0 |
|
| φ | 0.0161 (30) | 0.0 |
| φ | 0.0034 (39) | 0.0 |
|
| φ | 0.0023 (11) | 0.0 |
| φ | −0.0027 (17) | 0.07 |
|
| φ | 4.0522 (24) | 4.0369 |
| φ | 0.8951 (31) | 0.9163 |
|
|
| −1.41 (1) | 0.0 |
|
| −3.12 (14) | 0.0 |
|
|
| 226.5 (2) | 225 |
|
| −225.2 (2) | −225 |
|
|
| −0.13 (10) | 0.0 |
|
| −1.39 (12) | 0.0 |
|
| φ | 0.0073 (40) | 0.0 |
| φ | 0.0190 (50) | 0.0 |
|
| φ | 0.0004 (10) | 0.0 |
| φ | −0.0032 (10) | 0.0 |
|
| φ | 3.105 (48) | 3.1416 |
| φ | 0.0028 (31) | 0.0 |
|
|
| 178.15 (23) | 178.5 |
|
| −177.95 (20) | −178.5 |
|
|
| 138.49 (17) | 136.98 |
|
| −138.14 (13) | −136.98 |
|
|
| −0.51 (11) | 0.0 |
|
| −2.02 (10) | 0.0 |
|
| φ | 0.0187 (33) | 0.0 |
| φ | 0.0127 (36) | 0.0 |
|
| φ | 0.0038 (20) | 0.0 |
| φ | −0.0092 (11) | 0.0 |
|
| φ | 2.2322 (36) | 2.2253 |
| φ | −0.9098 (27) | −0.9163 |
|
Lattice parameters determined using the methods described within this paper
The values determined from cubic NaCl and monoclinic spodumene are shown. These values were determined both as a standard user of the existing understanding of the SXD instrument would determine them and using the new calibration of the instrument described in this paper. The results from the new SXDMEASURE method are shown and compared with the values determined using the HRPD instrument. Finally, a comparison is made for spodumene, where a restricted data set is used which restricts access to reciprocal space to ±10° from the horizontal direction on the detector surface. Where no value is given no value was refined.
| Data set |
|
|
| β (°) | σ | σ | σd |
|---|---|---|---|---|---|---|---|
| NaCl (90 reflections) | |||||||
| As engineered SXD | 5.6554 (59) | – | – | – | 0.75 | 0.35 | 0.0044 |
| Standard | 5.6519 (57) | – | – | – | 0.15 | 0.17 | 0.0020 |
| Calibrated | 5.6394 (55) | – | – | – | 0.16 | 0.17 | 0.0019 |
|
| 5.6403 (8) | – | – | – | 0.15 | 0.13 | 0.0013 |
| Spodumene (328 reflections) | |||||||
| As engineered SXD | 9.480 (9) | 8.476 (9) | 5.231 (4) | 110.28 (4) | 0.71 | 0.32 | 0.0037 |
| Standard | 9.479 (16) | 8.429 (14) | 5.232 (5) | 110.20 (8) | 0.15 | 0.19 | 0.0012 |
| Calibrated | 9.471 (9) | 8.394 (9) | 5.220 (4) | 110.21 (4) | 0.15 | 0.19 | 0.0019 |
|
| 9.459 (3) | 8.397 (4) | 5.218 (1) | 110.17 (3) | 0.14 | 0.14 | 0.0011 |
| HRPD | 9.46440 (16) | 8.38986 (6) | 5.21866 (42) | 110.1747 (12) | |||
| Restricted data set of spodumene (128 reflections in ±10°) | |||||||
|
| 9.476 (6) | 8.397 (5) | 5.218 (1) | 110.21 (2) | 0.14 | 0.18 | 0.0012 |
Demonstration of how the ability of an orientation matrix to predict the TOF for a given reflection is an important criterion in determining its accuracy
Here, the percentage of reflections for which the orientation matrix can predict the correct TOF within a given tolerance is shown, and it can be seen that, with increasing instrument description complexity, the ability to predict TOF correctly increases. dt/t is defined as the difference between the observed and calculated TOF for a given reflection (dt) divided by the TOF for the reflection.
| Calibration method | |d | |d | |d |
|---|---|---|---|
| As engineered | 33 | 54 | 81 |
| SXD normal | 45 | 74 | 94 |
| SXD calibrated | 52 | 84 | 98 |
|
| 67 | 90 | 99 |
A comparison of the ability of an orientation matrix to predict the position and time of reflections at detectors 1, 2 and 5
The orientation matrix was determined from data collected only at detectors 3 and 4 (as would happen at the beginning of a data collection) and used to predict reflection positions at detectors 1, 2, 5 and 6. Note that the normal SXD method does not allow calibration of the detector parameters at detectors 3 and 4 and hence is not included in this comparison. Here, dr is defined as dr = (dx 2 + dz 2)1/2, where dx and dz are the difference between the observed and calculated reflection positions on the detector surface. I is defined as .
| Calibration method | d | d | d | σ | σ | σd |
|---|---|---|---|---|---|---|
| As engineered | 0.94 | 5.7 | 10.3 | 0.76 | 0.44 | 0.0055 |
| SXD calibrated | 3.8 | 5.7 | 18.9 | 0.47 | 0.59 | 0.056 |
|
| 18 | 50 | 97 | 0.13 | 0.15 | 0.0009 |