| Literature DB >> 24902839 |
Håkan Emteborg1, Reinhard Zeleny2, Jean Charoud-Got2, Gustavo Martos2, Jörg Lüddeke3, Holger Schellin4, Katharina Teipel2.
Abstract
Coupling an infrared (IR) camera to a freeze dryer for on-line monitoring of freeze-drying cycles is described for the first time. Normally, product temperature is measured using a few invasive Pt-100 probes, resulting in poor spatial resolution. To overcome this, an IR camera was placed on a process-scale freeze dryer. Imaging took place every 120 s through a Germanium window comprising 30,000 measurement points obtained contact-free from -40 °C to 25 °C. Results are presented for an empty system, bulk drying of cheese slurry, and drying of 1 mL human serum in 150 vials. During freezing of the empty system, differences of more than 5 °C were measured on the shelf. Adding a tray to the empty system, a difference of more than 8 °C was observed. These temperature differences probably cause different ice structures affecting the drying speed during sublimation. A temperature difference of maximum 13 °C was observed in bulk mode during sublimation. When drying in vials, differences of more than 10 °C were observed. Gradually, the large temperature differences disappeared during secondary drying and products were transformed into uniformly dry cakes. The experimental data show that the IR camera is a highly versatile on-line monitoring tool for different kinds of freeze-drying processes.Entities:
Keywords: freeze-drying; human serum; infrared camera; infrared spetcroscopy; lyophilization; processing; spatial resolution; thermography; water in solids
Mesh:
Year: 2014 PMID: 24902839 PMCID: PMC4303947 DOI: 10.1002/jps.24017
Source DB: PubMed Journal: J Pharm Sci ISSN: 0022-3549 Impact factor: 3.534
Figure 1Integration of IR camera in freeze dryer. Top right shows the freeze-drying chamber in full and four shelves. Bottom left shows a magnification of area of interest with IR camera, viewing area, tray, and white holder for good alignment of the tray with the optics of the IR camera. During the operation of the freeze-drying programs, this holder had been removed.
Different Temperature Readouts at Different Time Intervals in the Freeze-Drying Program Using Probes, IR Camera, or Freeze-Dryer Control System
| Product and Figure Reference | Time (h) and Pressure | Product Temperature | Resistivity, Lyo RX (%) | IR Camera, Shelf/Tray/Product (°C), Average (Minimum/Maximum) | Temperature Difference (°C) | IR Camera Shelf | Freeze Dryer, Actual Shelf Temperature (°C) |
|---|---|---|---|---|---|---|---|
| Shelf ( | 0.25 (amb.), D | – | – | −1.8 | 5.1 | −1.4 | −2 |
| −4.9/0.2 | −1.7/−1.0 | ||||||
| Shelf ( | 0.75 (amb.), D | – | – | −38.5 | 5.7 | −39.8 | −37 |
| −41.8/−36.1 | −40.5/−39.0 | ||||||
| Shelf ( | 30 (0.2), S | – | – | −11.0 | 1.8 | −11.2 | −10 |
| −11.7/−10.1 | −11.6/−10.7 | ||||||
| Shelf | 68 (0.003), S | – | – | 19.5 | 0.9 | 19.6 | 20 |
| 19.0 / 19.9 | 19.3 / 20.0 | ||||||
| Tray ( | 0.25 (amb.), D | – | – | 2.6 | 8.2 | −0.8 | −3 |
| −2.6/5.6 | −1.2/−0.4 | ||||||
| Cheese ( | 1.4 (amb.), S | −34; −34; −30 | 74 | −10.3 | 29.8 | −36.2 | −40 |
| −29.5/0.3 | −36.8/−35.4 | ||||||
| Cheese ( | 30 (0.2), S | −13; −11; −10 | 98 | −9.8 | 13.0 | −11.6 | −11 |
| −18.7/−5.7 | −12.3/−11.2 | ||||||
| Cheese ( | 68 (0.006), S | 17; 15; 16 | 98 | 16.2 | 2.7 | 19.2 | 20 |
| 15.4/18.1 | 18.8/19.7 |
amb., ambient presssure; D, dynamic part of program; S, Static part of program.
Temperatures obtained from the probes as given in legend in Figure 3a.
Readout of average, maximum, and minimum in selected rectangles on masking tape placed on shelf in empty run using IRBIS®3, ε = 0.9. Average maximum and minimum was measured on the whole tray when loaded with cheese slurry, with ε = 0.95.
Temperature difference is referring to the difference of maximum and minimum values for the data column “shelf/tray/product.”
Readout of average, maximum, and minimum using IRBIS® 3 on masking tape denoted R1 (in Fig. 2a) placed next to freeze-drying tray, ε = 0.9.
Figure 3(a) Graphical output of the freeze-drying program for the cheese slurry as monitored by the freeze dryer. (b) Freeze drying program of human serum in vials as monitored by the freeze dryer. For both materials the traces for shelf temperature, product temperature and vacuum are most important.
Figure 2(a) Thermogram of empty run of shelf after 30 h of operation. Rectangle R1 shows the area of masking tape used to measure the shelf temperature using the IR camera in the presence of a tray. R2, R3, and R4 display maximum and minimum temperatures of each area, respectively, confirming an even temperature distribution on the shelf. For more information, cross-correlation with Table1 is necessary, ε = 0.90. (b) Thermogram of empty run of shelf after 15 min where temperature differences can be observed between inside walls and central parts of the channels where the silicon oil is flowing. Thin horizontal lines are narrow gaps between the stripes of masking tape. For more information, cross-correlation with Table1 is necessary, ε = 0.90. (c) Thermogram of empty run of shelf after 0.75 h operation at the end of cooling down where temperature differences can still be observed between inside walls and central parts of the channels. Thin horizontal lines are narrow gaps between the stripes of masking tape, ε = 0.90. (d) Thermogram of empty run with tray after 15 min. Top left corner is coldest, which suggests the best contact with the shelf, ε = 0.90. (e) Thermogram of run with cheese slurry after 1.4 h of operation in the freezing step. Central part is still not frozen and Pt-100 probe and Lyo control are visible to the left as light-green threads in the blue background, ε = 0.95. (f) Thermogram of run with cheese slurry after 30 h of operation in the sublimation step. Large temperature differences are apparent in different parts of the tray. Cracks in the slurry are also visible suggesting that certain areas are approaching dryness, ε = 0.95. (g) Thermogram of run with cheese slurry 68 h into the program in the secondary drying step; some water is still leaving the material because the temperature on the material is lower than the shelf. Pink/white horizontal bar on top left is rectangle of masking tape; R1 in Figure 2a used to monitor the shelf temperature, ε = 0.95.
Figure 4All the traces depicted in this figure represent temperature as a function of time with readings from 136 vials using the IR-camera every 2 min. Cross-correlation of product temperatures with Figure 3b is possible. (a) Freezing step of human serum. (b) Whole freeze-drying cycle of human serum. (c) Selected segment of the sublimation step of human serum.
Figure 5(a) Thermogram of 150 vials filled with human serum in the sublimation step after 20.9 h. The serum is visible as dark-blue spots. Occasionally, the vial is fitted with a Lyo-insert; in such cases, the vial is green and the temperature of the serum cannot be measured. The shelf temperature can easily be measured around each vial. (b) Thermogram of 70 vials filled with human serum in the sublimation step after 20.9 h with assigned measurement points. The serum is visible as dark-blue spots. (c) Thermogram of 150 vials filled with human serum in the secondary drying step after 36.6 h. The serum is visible as red spots. Occasionally, the vial is fitted with a Lyo-insert; in such cases, the vial is yellow and the the temperature of the serum cannot be measured.