Accuracy of surgical wound drainage measurements: an analysis and comparison.

Brian Yue1, Danielle Nizzero, Chunxiao Zhang, Natasha van Zyl, Jeannette Ting.   

Abstract

BACKGROUND: Surgical drain tube readings can influence the clinical management of the post-operative patient. The accuracy of these readings has not been documented in the current literature and this experimental study aims to address this paucity.
METHODS: Aliquots (10, 25, 40 and 90 mL) of black tea solution prepared to mimic haemoserous fluid were injected into UnoVac, RedoVac and Jackson-Pratt drain tubes. Nursing and medical staff from a tertiary hospital were asked to estimate drain volumes by direct observation; analysis of variance was performed on the results and significance level was set at 0.05.
RESULTS: Doctors and nurses are equally accurate in estimating drain tube volumes. Jackson-Pratt systems were found to be the most accurate for intermediate volumes of 25 and 40 mL. For extreme of volumes (both high and low), all drainage systems were inaccurate.
CONCLUSION: This study suggests that for intermediate volumes (25 and 40 mL), Jackson-Pratt is the drainage system of choice. The accuracy of volume measurement is diminished at the extremes of drain volumes; emptying of drainage systems is recommended to avoid overfilling of drainage systems.
© 2014 Royal Australasian College of Surgeons.

Entities:  

Keywords:  drain; post-operative; reading; reliability; suction

Mesh:

Year:  2014        PMID: 24891212     DOI: 10.1111/ans.12657

Source DB:  PubMed          Journal:  ANZ J Surg        ISSN: 1445-1433            Impact factor:   1.872


  1 in total

1.  How Much Blood Could a JP Suck If a JP Could Suck Blood?

Authors:  Erin M Mamuyac; Andrew K Pappa; Brian D Thorp; Charles S Ebert; Brent A Senior; Adam M Zanation; Feng-Chang Lin; Adam J Kimple
Journal:  Laryngoscope       Date:  2018-12-13       Impact factor: 3.325

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.