| Literature DB >> 24890996 |
Shinichiro Nakamura1, Adrija Sharma1, Kenji Nakamura2, Noboru Ikeda2, Sumesh M Zingde1, Richard D Komistek1.
Abstract
The purpose of the study was to analyze the effectiveness of an additional ball and socket articulation in implanted knees and whether it can replicate post-cam function. Fifteen knees implanted with a cruciate substituting (CS) polyethylene without a post and ten knees implanted with a posterior stabilized (PS) polyethylene with a post were analyzed using 3D model fitting approach. Two types of designs showed similar posterior translation and similar axial rotation. Most of the contact points at the ball and socket joint stayed within the socket height for the PS group. This study indicates that the ball and socket joint is able to function as a replacement of the post-cam mechanism, which might serve as a new way to achieve posterior stability.Entities:
Keywords: deep knee bend; knee kinematics; post-cam mechanism; total knee arthroplasty; tri-condylar implants
Mesh:
Substances:
Year: 2014 PMID: 24890996 DOI: 10.1016/j.arth.2014.04.024
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757