Literature DB >> 24887290

A novel single twist-drill access device for multimodal intracranial monitoring: a 5-year single-institution experience.

Muhammad Omar Chohan1, Omar S Akbik, Juan Ramos-Canseco, Pedro M Ramirez, Cristina Murray-Krezan, Tracey Berlin, Kim Olin, Christopher L Taylor, Howard Yonas.   

Abstract

BACKGROUND: Multimodal intracranial monitoring in the neurosurgical patient requires insertion of probes through multiple craniostomies.
OBJECTIVE: To report our 5-year experience with a novel device allowing multimodal monitoring though a single twist-drill hole.
METHODS: All devices (Hummingbird Synergy, Innerspace) were placed at the Kocher point between 2008 and 2013 at our institution. An independent clinical research nurse prospectively collected data on all bedside placements. Placement accuracy was graded on computed tomography scan as grade 1 (ipsilateral frontal horn or third ventricle), grade 2 (contralateral lateral ventricle), and grade 3 (anywhere else). Infection was monitored with serial cerebrospinal fluid samples.
RESULTS: Two hundred seventy-five devices (198 at bedside, 77 in operating room) were placed in patients with spontaneous subarachnoid hemorrhage (49%), traumatic brain injury (47%), and others (4%) for a median duration of 6 days. A junior (postgraduate year 1-2), midlevel (postgraduate year 3-4), or senior resident (postgraduate year 5-6) placed 39%, 32%, and 29% of the devices, respectively. Ninety-two percent of all devices placed were draining cerebrospinal fluid, ie, were grade 1 (75%) or 2 (17%). Placement accuracy did not vary with level of training. Complications included hemorrhage (10%) and infection (4%), with 1 patient requiring intraparenchymal hematoma evacuation and a second requiring abscess drainage. These rates were lower than reported in the literature for standard external ventricular drains.
CONCLUSION: Hummingbird Synergy is a novel single-port access device for multimodal intracranial monitoring that can be placed safely at the bedside or in the operating room with placement accuracy and has a complication profile similar to or better than that for standard external ventricular drains.

Entities:  

Mesh:

Year:  2014        PMID: 24887290     DOI: 10.1227/NEU.0000000000000451

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  3 in total

Review 1.  The Utility of Cerebral Blood Flow Assessment in TBI.

Authors:  Omar S Akbik; Andrew P Carlson; Mark Krasberg; Howard Yonas
Journal:  Curr Neurol Neurosci Rep       Date:  2016-08       Impact factor: 5.081

2.  Decision-making for decompressive craniectomy in traumatic brain injury aided by multimodality monitoring: illustrative case.

Authors:  Myranda B Robinson; Peter Shin; Robert Alunday; Chad Cole; Michel T Torbey; Andrew P Carlson
Journal:  J Neurosurg Case Lessons       Date:  2021-06-21

3.  Comparison of parenchymal and ventricular intracranial pressure readings utilizing a novel multi-parameter intracranial access system.

Authors:  Tracey Berlin; Cristina Murray-Krezan; Howard Yonas
Journal:  Springerplus       Date:  2015-01-06
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.