Literature DB >> 24878237

Percutaneous vascular closure using an anchored collagen plug provides effective haemostasis following both antegrade and retrograde femoral arterial punctures.

A Chaudhuri1, A York2, R Dey3.   

Abstract

OBJECTIVES: Small published series suggest a higher failure rate for Angio-Seal vascular closure device (VCD) deployment after antegrade femoral puncture, despite the need for shorter haemostasis times, early discharge, and possibly higher turnover. We seek to compare the deployment efficacy and complications of the Angio-Seal VCD between antegrade and retrograde femoral arterial deployments.
METHODS: Radiological data was retrospectively analysed from prospective databases from the hospitals' Computerised Radiology Information System (CRIS) over 2010-2012. Angio-Seal gauge, Rutherford class (as applicable), puncture mode (used to classify deployment as antegrade/retrograde), sheath sizes, and deployment success/failures were recorded. Numerical/statistical analyses were undertaken using Microsoft Excel 10/SISA software.
RESULTS: A total of 519 Angio-Seal VIP VCDs were deployed in 470 patients over 2010-2012 (13 other patients could not be analysed due to incomplete data). Sheath sizes for antegrade/retrograde femoral puncture were 5F, n = 22/9; 6F, n = 244/223; 7F, n = 1/5; 9F, n = 4/0. 8F Angio-Seal VIPs were used for 9F punctures only, 6F for the remainder. The overall deployment success rate was 93.7%. In total, 247 (91.1%) successful antegrade deployments were undertaken with 24 (8.9%) failures, compared with 229 (96.6%) successful retrograde deployments with eight (3.4%) failures. Antegrade/retrograde failures were classed as failure to deploy, n = 15/5; bleeding despite successful deployment requiring supplementary compression, n = 6/1; haematoma formation, n = 2/1; groin pain, n = 0/1; vessel stenosis, n = 1/0. Higher deployment failures were noted with antegrade deployment (p < .02, chi-square test).
CONCLUSIONS: Angio-Seal deployment is successful for both antegrade/retrograde femoral punctures albeit with a higher antegrade failure rate.
Copyright © 2014 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Angio-Seal; Angiography/angioplasty; Femoral artery puncture; Hemostasis; Vascular closure device

Mesh:

Substances:

Year:  2014        PMID: 24878237     DOI: 10.1016/j.ejvs.2014.04.017

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  1 in total

1.  Successful Retrieval of an Embolized Vascular Closure Device (Angio-Seal®) After Peripheral Angioplasty.

Authors:  Philipp Jud; Rupert Portugaller; Dennis Bohlsen; Thomas Gary; Marianne Brodmann; Gerald Hackl; Franz Hafner
Journal:  Cardiovasc Intervent Radiol       Date:  2017-01-18       Impact factor: 2.740