Literature DB >> 24876218

Delayed fracture of MatrixRIB precontoured plate system.

Calvin S H Ng1, Randolph H L Wong2, Micky W T Kwok2, Anthony P C Yim2.   

Abstract

OBJECTIVES: Use of titanium prostheses for reconstructing chest wall defects following major chest wall resections is rapidly increasing in popularity. Previously, complications including prosthesis fracture have been reported for the system secured to the rib ends using clips following chest wall reconstruction and pectus excavatum repair. By contrast, fracture failure in titanium systems fastening the plate to the rib with locking screws through predrilled holes has not been previously reported, possibly owing to differences in the design and material of the system.
METHODS: We report an index case of plate fracture in the latter design following rib reconstruction and discuss the pathophysiology and relative risks behind such fractures.
RESULTS: Fracture of titanium plate occurred after chest impact during sport at 25 months following initial plate implantation. Surgical excision was performed because of persistent pain and cough.
CONCLUSIONS: Fracture failure can occur in those systems fastening the plate to the rib with locking-screws, and patients, particularly those engaged in contact sports or occupations with chest impact risks, should be informed of this possibility during consent, as surgical removal is most likely required following implant fracture failur.
© The Author 2014. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.

Entities:  

Keywords:  Chest wall; Fracture; MatrixRIB; Reconstruction; Titanium plate

Mesh:

Substances:

Year:  2014        PMID: 24876218     DOI: 10.1093/icvts/ivu175

Source DB:  PubMed          Journal:  Interact Cardiovasc Thorac Surg        ISSN: 1569-9285


  5 in total

Review 1.  Multi-dimensional printing in thoracic surgery: current and future applications.

Authors:  Jackson K S Kwok; Rainbow W H Lau; Ze-Rui Zhao; Peter S Y Yu; Jacky Y K Ho; Simon C Y Chow; Innes Y P Wan; Calvin S H Ng
Journal:  J Thorac Dis       Date:  2018-04       Impact factor: 2.895

2.  Ten-Year Experience of Chest Wall Reconstruction: Retrospective Review of a Titanium Plate MatrixRIB™ System.

Authors:  Teddy H Y Wong; Ivan C H Siu; Kareem K N Lo; Ethan Y H Tsang; Innes Y P Wan; Rainbow W H Lau; T W Chiu; Calvin S H Ng
Journal:  Front Surg       Date:  2022-07-05

Review 3.  Recent clinical innovations in thoracic surgery in Hong Kong.

Authors:  Ze-Rui Zhao; Zheng Li; Dong-Rong Situ; Calvin S H Ng
Journal:  J Thorac Dis       Date:  2016-08       Impact factor: 2.895

4.  Chest wall stabilization and reconstruction: short and long-term results 5 years after the introduction of a new titanium plates system.

Authors:  Angela De Palma; Francesco Sollitto; Domenico Loizzi; Francesco Di Gennaro; Daniele Scarascia; Annalisa Carlucci; Giuseppe Giudice; Andrea Armenio; Rossana Ludovico; Michele Loizzi
Journal:  J Thorac Dis       Date:  2016-03       Impact factor: 2.895

5.  An unexpected fracture of a titanium rib plate in a patient with traumatic brain injury.

Authors:  Ki Tae Kim; Sung Wook Chang
Journal:  Trauma Case Rep       Date:  2019-10-31