| Literature DB >> 24875995 |
Jenq-Nan Yih, Yvonne Yuling Hu, Yong Da Sie, Li-Chung Cheng, Chi-Hsiang Lien, Shean-Jen Chen.
Abstract
This Letter presents an enhanced temporal focusing-based multiphoton excitation (MPE) microscope in which the conventional diffraction grating is replaced by a digital micromirror device (DMD). Experimental results from imaging a thin fluorescence film show that the 4.0 μm axial resolution of the microscope is comparable with that of a setup incorporating a 600 lines/mm grating; hence, the optical sectioning ability of the proposed setup is demonstrated. Similar to a grating, the DMD diffracts illuminating light frequencies for temporal focusing; additionally, it generates arbitrary patterns. Since the DMD is placed on the image-conjugate plane of the objective lens' focal plane, the MPE pattern can be projected on the focal plane precisely.Entities:
Mesh:
Year: 2014 PMID: 24875995 DOI: 10.1364/OL.39.003134
Source DB: PubMed Journal: Opt Lett ISSN: 0146-9592 Impact factor: 3.776