Abbott ARCHITECT iPhenytoin assay versus similar assays for measuring free phenytoin concentrations.

Danyel Hermes Tacker, Randy Robinson, Peter L Perrotta.   

Abstract

OBJECTIVE: To measure free phenytoin (FP) concentrations in filtered specimens using the Abbott ARCHITECT iPhenytoin assay and to compare results from this method with results from the Abbott TDx/FLx assays.
METHODS: We verified accuracy, analytic measurement range, and precision for FP measurements. For correlation and therapeutic interval studies, we used filtered calibrators, controls, proficiency-testing materials, and surplus clinical samples. After implementation, we determined proficiency testing results.
RESULTS: The analytic measurement range was 2.0 to 25.0 micromol/L. Quality control materials (6.1, 12.6, and 20.1 micromol/L) provided mean (SD) recoveries of 96.1 (5.0%), 99.2 (5.0%), and 99.3 (5.7%), respectively, and coefficients of variation of 5.2%, 5.0%, and 5.8%, respectively. Clinical specimens produced mean (SD) FP recovery levels of 103.7 (10.6%) (bias, 0.1 [0.3] micromol/L). Altering the FP therapeutic range (4.0-8.0 micromol/L) was unnecessary. Proficiency testing yielded consistently acceptable results.
CONCLUSION: Our accuracy, precision, and correlation results were similar for the TDx/FLx and ARCHITECT assays, which demonstrates that the ARCHITECT iPhenytoin assay is acceptable for clinical FP measurements.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24869001     DOI: 10.1309/lm28b9dsrjcbcwwj

Source DB:  PubMed          Journal:  Lab Med        ISSN: 0007-5027


  1 in total

1.  Estimation of Free Phenytoin Concentration in Critically Ill Patients with Hypoalbuminemia: Direct-measurement vs Traditional Equations.

Authors:  Premila M Wilfred; Sumith Mathew; Binila Chacko; Ratna Prabha; Binu Susan Mathew
Journal:  Indian J Crit Care Med       Date:  2022-06
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.