Literature DB >> 2486368

An in vitro comparative study of St. Jude Medical and Edwards-Duromedics bileaflet valves using laser anemometry.

R Fatemi1, K B Chandran.   

Abstract

An in vitro comparative study of St. Jude (SJ) and Edwards-Duromedics (DM) Bileaflet valves was performed under steady and physiological pulsatile flow conditions in an axisymmetric chamber using Laser Doppler Anemometry (LDA). LDA measurements were conducted in two different orientations; in the first orientation, the LDA traverse was perpendicular and, in the second orientation, parallel to the tilt axis of the leaflets. The axial velocities were measured in both orientations at two different locations distal to the valves. The velocity profiles at peak systole show the presence of stronger vortex in the sinus region for flow past SJ valve in the first orientation compared to the DM valve. Velocity profile distal to the SJ valve in second orientation was relatively flat where as for the DM valve, a jet-like flow was present. The differences found in the velocity profiles between the two valves can be attributed to the differences in geometry with thicker leaflets, smaller angle of leaflets opening and the presence of the leaflet curvature for the DM valve. The results obtained in this study do not show any fluid dynamic advantages due to the curved leaflet geometry of the DM valve.

Entities:  

Mesh:

Year:  1989        PMID: 2486368     DOI: 10.1115/1.3168382

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  3 in total

1.  Dynamics of a mechanical monoleaflet heart valve prosthesis in the closing phase: effect of squeeze film.

Authors:  C Gill-Jeong; K B Chandran
Journal:  Ann Biomed Eng       Date:  1995 Mar-Apr       Impact factor: 3.934

Review 2.  Simulation of Mechanical Heart Valve Dysfunction and the Non-Newtonian Blood Model Approach.

Authors:  Aolin Chen; Adi Azriff Bin Basri; Norzian Bin Ismail; Masaaki Tamagawa; Di Zhu; Kamarul Arifin Ahmad
Journal:  Appl Bionics Biomech       Date:  2022-04-19       Impact factor: 1.664

3.  Adverse Hemodynamic Conditions Associated with Mechanical Heart Valve Leaflet Immobility.

Authors:  Fardin Khalili; Peshala P T Gamage; Richard H Sandler; Hansen A Mansy
Journal:  Bioengineering (Basel)       Date:  2018-09-16
  3 in total