| Literature DB >> 2486366 |
Abstract
This paper describes the production and calibration of a miniature psychrometer treated with a specially developed porous coating. The investigation was conducted to determine localized patterns of rapidly changing temperature and relative humidity in dynamic flowing gas environments (e.g., with particular attention to future applications to the human respiratory system). The technique involved the use of dry miniature thermocouples and wetted miniature thermocouples coated with boron nitride to act as a wicking material. A precision humidity generator was developed for calibrating the psychrometer. It was found that, in most cases, the measured and expected (i.e., theoretically predicted) relative humidity agreed to within 0.5 to 1.0 percent relative humidity. Procedures that would decrease this discrepancy even further were pinpointed, and advantages of using the miniature psychrometer were assessed.Entities:
Mesh:
Year: 1989 PMID: 2486366 DOI: 10.1115/1.3168380
Source DB: PubMed Journal: J Biomech Eng ISSN: 0148-0731 Impact factor: 2.097