Design and development of a micro-thermocouple sensor for determining temperature and relative humidity patterns within an airstream.

A D Eisner1, T B Martonen.   

Abstract

This paper describes the production and calibration of a miniature psychrometer treated with a specially developed porous coating. The investigation was conducted to determine localized patterns of rapidly changing temperature and relative humidity in dynamic flowing gas environments (e.g., with particular attention to future applications to the human respiratory system). The technique involved the use of dry miniature thermocouples and wetted miniature thermocouples coated with boron nitride to act as a wicking material. A precision humidity generator was developed for calibrating the psychrometer. It was found that, in most cases, the measured and expected (i.e., theoretically predicted) relative humidity agreed to within 0.5 to 1.0 percent relative humidity. Procedures that would decrease this discrepancy even further were pinpointed, and advantages of using the miniature psychrometer were assessed.

Entities:  

Mesh:

Year:  1989        PMID: 2486366     DOI: 10.1115/1.3168380

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  1 in total

Review 1.  Behavior of hygroscopic pharmaceutical aerosols and the influence of hydrophobic additives.

Authors:  A J Hickey; T B Martonen
Journal:  Pharm Res       Date:  1993-01       Impact factor: 4.200