Safety and efficacy of different paclitaxel-eluting balloons in a porcine model.

Armando Pérez de Prado1, Claudia Pérez-Martínez2, Carlos Cuellas Ramón2, Marta Regueiro Purriños2, Alejandro Diego Nieto2, José M Gonzalo-Orden2, María Molina Crisol3, Alex Gómez Castel3, Luis Duocastella Codina3, Felipe Fernández-Vázquez2.   

Abstract

INTRODUCTION AND
OBJECTIVES: Paclitaxel-eluting balloons have shown high antiproliferative efficacy in the treatment and prevention of restenosis. Nevertheless, not all available devices are equally effective, which makes it interesting to compare results in a preclinical model. Our objective was to assess the preclinical efficacy and safety of different devices.
METHODS: We implanted 51 metallic stents (Architect(®), iVascular) in 17 domestic swine (mean, 25 [3] kg), inserting 1 stent per major coronary artery. Stent postdilatation was performed with different control balloons (n=10) or paclitaxel-eluting balloons: paclitaxel-eluting balloon 1 (iVascular) (n=15); paclitaxel-eluting balloon 2 (iVascular) (n=16) and In.Pact Falcon(®) (Medtronic) (n=10). The restenosis rate (using angiography and histomorphometry) and vascular healing parameters (balloon-related vascular injury score, endothelialization rate, and fibrin and inflammation scores) were analyzed at 28 days.
RESULTS: The distinct paclitaxel-eluting balloons showed a similar degree of stenosis at follow-up, which was significantly lower than that in the control group: diameter stenosis was 9% (12%) vs 34% (18%) by angiography (P<.0001) and was 22% (8%) vs 51% (18%) by histomorphometry (P<.0001). Scores for vascular injury (mean, 0.6 [0.5]) and inflammation (mean, 0.8 [0.3]) were uniformly low across all groups. Drug effect markers differed significantly between the paclitaxel-eluting balloons and control groups, with lower endothelialization rates (87% [10%] vs 99% [2%]; P=.0007) and higher fibrin scores (2.1 [0.7] vs 0.4 [0.5]; P<.0001) in the paclitaxel-eluting balloons groups. There were no differences between the different paclitaxel-eluting balloons.
CONCLUSIONS: In this preclinical model, the paclitaxel-eluting balloons studied significantly reduced in-stent restenosis compared with the control balloons. Although there were no findings of persistent vascular injury or inflammation, delayed endothelialization and fibrin aggregate suggest a drug deposition response.
Copyright © 2013 Sociedad Española de Cardiología. Published by Elsevier Espana. All rights reserved.

Entities:  

Keywords:  Animal model; Balón liberador de paclitaxel; Modelo animal; Paclitaxel-eluting balloon; Reparación vascular; Restenosis; Stent; Vascular healing

Mesh:

Substances:

Year:  2014        PMID: 24863594     DOI: 10.1016/j.rec.2013.09.030

Source DB:  PubMed          Journal:  Rev Esp Cardiol (Engl Ed)        ISSN: 1885-5857


  3 in total

1.  Impact of drug-eluting balloon (pre- or post-) dilation on neointima formation in de novo lesions treated by bare-metal stent: the IN-PACT CORO trial.

Authors:  Francesco Burzotta; Marta Francesca Brancati; Carlo Trani; Giancarlo Pirozzolo; Gianluigi De Maria; Antonio Maria Leone; Giampaolo Niccoli; Italo Porto; Francesco Prati; Filippo Crea
Journal:  Heart Vessels       Date:  2015-04-12       Impact factor: 2.037

2.  Reduced histologic neo in-stent restenosis after use of a paclitaxel-coated cutting balloon in porcine coronary arteries.

Authors:  Denise Traxler; Rayyan Hemetsberger; Andreas Spannbauer; Katrin Zlabinger; Alfred Gugerell; Dominika Lukovic; Ljubica Mandic; Noemi Pavo; Johannes Winkler; Mariann Gyöngyösi
Journal:  Histol Histopathol       Date:  2019-10-24       Impact factor: 2.303

3.  Association of serum apoA-I with in-stent restenosis in coronary heart disease.

Authors:  Xin Wang; Min Zhang; Jie Cheng; Hua Zhou
Journal:  BMC Cardiovasc Disord       Date:  2022-08-04       Impact factor: 2.174

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.