PURPOSE: Clinical investigational studies were conducted to demonstrate the accuracy and reproducibility of the Illumina MiSeqDx CF System, a next-generation sequencing (NGS) in vitro diagnostic device for cystic fibrosis testing. METHODS: Two NGS assays - a Clinical Sequencing Assay (Sequencing Assay) and a 139-Variant Assay (Variant Assay) - were evaluated in both an Accuracy Study and a Reproducibility Study, with comparison to bi-directional Sanger sequencing and PCR as reference methods. For each study, positive agreement (PA), negative agreement (NA), and overall agreement (OA) were evaluated. RESULTS: In the Accuracy Study, the Sequencing Assay achieved PA of 99.7% including the polyTG/polyT region and PA of 100% excluding the region. The Variant Assay achieved PA of 100%. NA and OA were >99.99% for both Assays. In the Reproducibility Study, the Sequencing Assay achieved PA of 99.2%; NA and OA were both 99.7%. The Variant Assay achieved PA of 99.8%; NA and OA were both 99.9%. Sample pass rates were 99.7% in both studies for both assays. CONCLUSION: This is the first systematic evaluation of a NGS platform for broad clinical use as an in vitro diagnostic, including accuracy validation with multiple reference methods and reproducibility validation at multiple clinical sites. These NGS-based Assays had accurate and reproducible results which were comparable to or better than other methods currently in clinical use for clinical genetic testing of cystic fibrosis.
PURPOSE: Clinical investigational studies were conducted to demonstrate the accuracy and reproducibility of the Illumina MiSeqDx CF System, a next-generation sequencing (NGS) in vitro diagnostic device for cystic fibrosis testing. METHODS: Two NGS assays - a Clinical Sequencing Assay (Sequencing Assay) and a 139-Variant Assay (Variant Assay) - were evaluated in both an Accuracy Study and a Reproducibility Study, with comparison to bi-directional Sanger sequencing and PCR as reference methods. For each study, positive agreement (PA), negative agreement (NA), and overall agreement (OA) were evaluated. RESULTS: In the Accuracy Study, the Sequencing Assay achieved PA of 99.7% including the polyTG/polyT region and PA of 100% excluding the region. The Variant Assay achieved PA of 100%. NA and OA were >99.99% for both Assays. In the Reproducibility Study, the Sequencing Assay achieved PA of 99.2%; NA and OA were both 99.7%. The Variant Assay achieved PA of 99.8%; NA and OA were both 99.9%. Sample pass rates were 99.7% in both studies for both assays. CONCLUSION: This is the first systematic evaluation of a NGS platform for broad clinical use as an in vitro diagnostic, including accuracy validation with multiple reference methods and reproducibility validation at multiple clinical sites. These NGS-based Assays had accurate and reproducible results which were comparable to or better than other methods currently in clinical use for clinical genetic testing of cystic fibrosis.
Authors: Martina I Lefterova; Peidong Shen; Justin I Odegaard; Eula Fung; Tsoyu Chiang; Gang Peng; Ronald W Davis; Wenyi Wang; Martin Kharrazi; Iris Schrijver; Curt Scharfe Journal: J Mol Diagn Date: 2016-02-01 Impact factor: 5.568
Authors: Daniel Trujillano; Maximilian E R Weiss; Julia Köster; Efstathios B Papachristos; Martin Werber; Krishna Kumar Kandaswamy; Anett Marais; Sabrina Eichler; Jenny Creed; Erol Baysal; Iqbal Yousuf Jaber; Dina Ahmed Mehaney; Chantal Farra; Arndt Rolfs Journal: Mol Genet Genomic Med Date: 2015-04-16 Impact factor: 2.183