Literature DB >> 24841731

Magnetic systems for tarsorrhaphy.

Anne Barmettler1, Nadee Nissanka, Mark I Rosenblatt, Rohini Rao, David Lipson, Gary J Lelli.   

Abstract

PURPOSE: Paralytic lagophthalmos can cause exposure keratopathy. Current treatments have difficulties: extrusion, migration, allergic reaction, anatomical disruption, and technically difficult surgeries. The goal of this study was to design, create, and assess a new method for eyelid closure using magnets.
METHODS: This experimental study is approved by institutional review board. Creation of eyelid model and magnet systems: A model of the eyelid and eyeball was constructed to determine the necessary magnetic strength. Neodymium magnets were cast in silicone mold carriers. Assessment of temporary magnet systems in humans: 1) a magnet was affixed to the upper and lower eyelids, and 2) a magnet was affixed to the upper eyelid and another to eyeglasses. Parameters evaluated were eyelid positions and success of eyelid closure in healthy adult volunteers.
RESULTS: Magnetic force required to create the average eyelid opening force was calculated from the eyelid model to be equivalent to 4 magnets. The magnet system affixed to upper and lower eyelids resulted in complete eyelid closure in 5 of 5 normal controls, while the magnet system affixed to the upper eyelid and to spectacles resulted in complete eyelid closure in 10 of 13 normal controls.
CONCLUSIONS: Magnetic systems for eyelid closure were designed that used either magnets affixed to both upper and lower eyelids or an upper eyelid magnetic component combined with a magnetic element in the lower rim of spectacles. Both were effective in eyelid closure in a model eye and normal controls. These systems may ultimately provide a simplified, safer, and less invasive method to treat paralytic lagophthalmos.

Entities:  

Mesh:

Year:  2014        PMID: 24841731     DOI: 10.1097/IOP.0000000000000097

Source DB:  PubMed          Journal:  Ophthalmic Plast Reconstr Surg        ISSN: 0740-9303            Impact factor:   1.746


  3 in total

1.  Putting the Pieces Back Together: Optimizing Function and Appearance after Orbital Surgery.

Authors:  Kimberly Cockerham; Jacquelyn Laplant
Journal:  J Neurol Surg B Skull Base       Date:  2021-02-22

2.  A Prototype External Magnetic Eyelid Device for Blepharoptosis.

Authors:  Kevin E Houston; Matteo Tomasi; Michael Yoon; Eleftherios I Paschalis
Journal:  Transl Vis Sci Technol       Date:  2014-12-17       Impact factor: 3.283

Review 3.  The use of neodymium magnets in healthcare and their effects on health.

Authors:  Cengiz Yuksel; Seyit Ankarali; Nehir Aslan Yuksel
Journal:  North Clin Istanb       Date:  2018-09