Colon high-resolution manometry: using pressure topography plots to evaluate pediatric colon motility.

Khalil I El-Chammas1, Neelesh A Tipnis, Pippa M Simpson, Manu R Sood.   

Abstract

OBJECTIVES: Colon manometry is usually performed using the 8-pressure sensor water-perfused manometry system. High-resolution manometry (HRM), using closely spaced solid-state pressure recording sensors, provides more detailed information of gut luminal pressure changes, and, by displaying the HRM data as a pressure topography plot (PTP), helps with data interpretation. Our aim was to compare the colon and rectal luminal pressure data obtained using 8 pressure sensors and displayed as conventional line plot (CLP) with data obtained using a custom-made solid state manometry catheter with 36 pressure recording sensors and displayed as PTP.
METHODS: We evaluated colon manometry patterns during fasting, response to meal, and bisacodyl stimulation in 10 patients with constipation and stool expulsion disorders. The data from 8 pressure sensors were displayed as CLP and data from 36 pressure sensors as PTP. Two gastroenterologists independently interpreted these studies. We calculated variability in interpreting colon, rectal, and anal manometry data.
RESULTS: Intermode, interobserver, and intraobserver reliability were good to excellent for recognizing colon contraction patterns when data are displayed as PTP compared with when displayed as CLP, whereas the reliability for recognizing anal contractions were poor to excellent.
CONCLUSIONS: Colonic and anal manometry patterns are easily recognized when HRM data are expressed as PTP. Obtaining information of colonic luminal pressure changes with rectum and anal pressure changes using HRM can help better understand the pathophysiology of pediatric constipation and stool expulsion disorders.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24840515     DOI: 10.1097/MPG.0000000000000442

Source DB:  PubMed          Journal:  J Pediatr Gastroenterol Nutr        ISSN: 0277-2116            Impact factor:   2.839


  3 in total

Review 1.  High-resolution colonic manometry and its clinical application in patients with colonic dysmotility: A review.

Authors:  Yu-Wei Li; Yong-Jun Yu; Fei Fei; Min-Ying Zheng; Shi-Wu Zhang
Journal:  World J Clin Cases       Date:  2019-09-26       Impact factor: 1.337

Review 2.  First translational consensus on terminology and definitions of colonic motility in animals and humans studied by manometric and other techniques.

Authors:  Maura Corsetti; Marcello Costa; Gabrio Bassotti; Adil E Bharucha; Osvaldo Borrelli; Phil Dinning; Carlo Di Lorenzo; Jan D Huizinga; Marcel Jimenez; Satish Rao; Robin Spiller; Nick J Spencer; Roger Lentle; Jasper Pannemans; Alexander Thys; Marc Benninga; Jan Tack
Journal:  Nat Rev Gastroenterol Hepatol       Date:  2019-07-11       Impact factor: 46.802

3.  Simultaneous assessment of colon motility in children with functional constipation by cine-MRI and colonic manometry: a feasibility study.

Authors:  M H Vriesman; C S de Jonge; S Kuizenga-Wessel; B Adler; A Menys; A J Nederveen; J Stoker; M A Benninga; C Di Lorenzo
Journal:  Eur Radiol Exp       Date:  2021-02-10
  3 in total