Felix Grases1, Adrian Rodriguez1, Francisco Berga1, Antonia Costa-Bauza1, Rafael Maria Prieto1, Isabel Burdallo2, Alfredo Cadarso2, Cecilia Jimenez-Jorquera2, Antonio Baldi2, Rosendo Garganta3.
Abstract
PURPOSE: Urinary pH is an important factor linked to renal stone disease and a useful marker in the treatment of urolithiasis. Although the gold standard for measuring urinary pH utilizes a glass electrode and a pH meter, at present dipstick testing is largely used to estimate urinary pH. However, the accuracy and precision of this method may be insufficient for making clinical decisions in patients with lithiasis. The aim of this study is to describe a new device for urinary pH testing.Entities:
Keywords: Renal lithiasis; Urinary pH; pH measurement; pH self-control
Year: 2014 PMID: 24839588 PMCID: PMC4022968 DOI: 10.1186/2193-1801-3-209
Source DB: PubMed Journal: Springerplus ISSN: 2193-1801
Figure 1Images of the device (left), and pH sensor probe (right).
Figure 2Scheme of the experimental set-up.
Mean and standard deviation of pH values of two buffer solutions measured with the new electronic device
| 0.7 M acetate buffer pH = 5.5 | 0.15 M phosphate buffer pH = 6.5 | |||
|---|---|---|---|---|
| Mean | Standard deviation | Mean | Standard deviation | |
| Intra-day (n = 5) | 5.42 | 0.04 | 6.50 | 0.00 |
| Inter-day (n = 5) | 5.46 | 0.05 | 6.50 | 0.00 |
Figure 3Scatter graphs showing the correlation of pH measurements between pH meter and the dipstick (3a) and the new device (3b) in fresh urine. The regression line is illustrated by a solid line, while the unity line is shown in dotted.
pH values measured with the dipstick and the pH-meter in fresh urine
| pH dipstick | pH-meter Mean (±SD) | Difference (±SD) | n |
|---|---|---|---|
| 5,0 | 5,14 ± 0.07 | 0,14 ± 0.07 | 5 |
| 5,5 | 5,34 ± 0.17 | 0,16 ± 0.17 | 10 |
| 6,0 | 5,78 ± 0.18 | 0,22 ± 0.18 | 6 |
| 6,5 | 6,33 ± 0.16 | 0,17 ± 0.16 | 7 |
| 7,0 | 0 | ||
| 7,5 | 7,44 ± 0.07 | 0,06 ± 0.07 | 2 |
pH values measured with the new electronic device and the pH-meter in fresh urine
| pH interval | pH-meter Mean (±SD) | New device Mean (±SD) | Difference (±SD) | n |
|---|---|---|---|---|
| 5-5.5 | 5,17 ± 0.08 | 5,20 ± 0.12 | 0.03 ± 0.14 | 10 |
| 5.5-6 | 5,60 ± 0.14 | 5,68 ± 0.14 | 0.08 ± 0.20 | 10 |
| 6-6.5 | 6,26 ± 0.14 | 6,29 ± 0.12 | 0.03 ± 0.18 | 7 |
| 6.5-7.5 | 7,17 ± 0.50 | 7,27 ± 0.49 | 0.10 ± 0.70 | 3 |