A simulator study of tube exchange with three different designs of double-lumen tubes.

Ryan Gamez1, Peter Slinger.   

Abstract

BACKGROUND: We sought to determine whether the design of 3 different double-lumen endobronchial tubes (DLT) (Rusch, Mallinckrodt, Fuji) has an effect on the ease of placement over an airway exchange catheter (AEC) using a video laryngoscope.
METHODS: A convenience sample of 17 anesthesia residents and fellows with at least 3 years of anesthesia training was recruited from teaching hospitals in Toronto for a randomized crossover trial. Each participant passed each DLT over an AEC in an airway simulator, visualized and video recorded via a video laryngoscope (GlideScope). The order of exchange was randomized by blindly pulling the name of the manufacturer of a DLT from a box. The primary outcome was time to intubate, defined as time from the bronchial lumen entering the GlideScope view to the bronchial lumen passing the vocal cords. Also recorded were participants' subjective rating of the ease of use and failure rate, defined as an attempt >150-second duration.
RESULTS: Time to intubate was faster with the Fuji-Phycon DLT (median 2 seconds) compared with both the Rusch (median 27 seconds, P = 0.0144) and Mallinckrodt (median 21 seconds, P = 0.0117). On a scale of 1 to 10, with 10 being very easy to use and 1 being very difficult, the Fuji-Phycon was judged to be easier to use (median 10 seconds) compared with the Rusch (median 3, P = 0.0186) and the Mallinckrodt (median 4 seconds, P = 0.0123). The Rusch was associated with significantly more failures than the other DLTs, P = 0.002.
CONCLUSIONS: The Fuji-Phycon DLT was easier to pass over an AEC in this simulator trial and warrants consideration in patients with difficult airways who require 1-lung ventilation.

Entities:  

Mesh:

Year:  2014        PMID: 24836474     DOI: 10.1213/ANE.0000000000000250

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  4 in total

1.  Bronchial lumen is the safer route for an airway exchange catheter in double-lumen tube replacement: preventable complication in airway management for thoracic surgery.

Authors:  Hsiang-Ling Wu; Ying-Hsuan Tai; Ling-Fang Wei; Hung-Wei Cheng; Chiu-Ming Ho
Journal:  J Thorac Dis       Date:  2017-10       Impact factor: 2.895

2.  Time to tracheal intubation over a fibreoptic bronchoscope using a silicone left double-lumen endobronchial tube versus polyvinyl chloride single-lumen tube with bronchial blocker: a randomized controlled non-inferiority trial.

Authors:  Ji Young Yoo; Yun Jeong Chae; Sung Yong Park; Seokjin Haam; Myungseob Kim; Dae Hee Kim
Journal:  J Thorac Dis       Date:  2019-03       Impact factor: 2.895

Review 3.  Videolaryngoscopes for placement of double lumen tubes: Is it time to say goodbye to direct view?

Authors:  M R El-Tahan
Journal:  Saudi J Anaesth       Date:  2016 Apr-Jun

4.  Appropriate tube temperature for fiberoptic bronchoscope-guided intubation of thermally softened double-lumen endotracheal tubes: A CONSORT-compliant article.

Authors:  Yang Yu; Qianqian Jia; Lijie Zhou; Zhou Liu; Shujuan Liang; Zhen Yang; Qiong Wan
Journal:  Medicine (Baltimore)       Date:  2022-10-07       Impact factor: 1.817

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.