Repeatability of grading meibomian gland dropout using two infrared systems.

William Ngo1, Sruthi Srinivasan, Marc Schulze, Lyndon Jones.   

Abstract

PURPOSE: To determine the interobserver and intraobserver repeatability in using the OCULUS Keratograph 4 (K4) and 5M (K5M) to grade meibomian gland (MG) dropout using meibography grading scales.
METHODS: The inferior and superior eyelids of 40 participants (35 women, 5 men; mean age = 32 years) were imaged three times each on both instruments. The images were split into one training and two study sets; the latter were graded (four-point meibography scale) by two observers on two separate occasions (24 hours apart) to determine repeatability. Semiobjective quantification of percentage MG dropout was conducted using ImageJ on K4 and K5M images. A finer seven-point meibography scale was used to grade a separate set of K5M images.
RESULTS: For the four-point scale, interobserver mean difference (MD) (±SD) was 0.08 (±0.55) on day 1 and 0.13 (±0.50) on day 2, and the concordance correlation coefficient (CCC) was 0.79 and 0.81 on days 1 and 2, respectively. Intraobserver MD (±SD) was 0.04 (±0.54), CCC = 0.79 for observer 1; intraobserver MD (±SD) was -0.09 (±0.60), CCC = 0.74 for observer 2. For the seven-point scale, interobserver MD (±SD) was 0.05 (±0.45), CCC = 0.89 on day 1, and interobserver MD (±SD) was 0.01 (±0.41), CCC = 0.91 on day 2. Intraobserver MD (±SD) was -0.10 (±0.35), CCC = 0.93 for observer 1, and intraobserver MD (±SD) was -0.06 (±0.30), CCC = 0.95 for observer 2. Percentage dropout measured between the K4 and K5M images showed lack of agreement, with 21.8% coefficient of repeatability. There was no significant correlation (r < 0.2; p > 0.05) between meibography score and clinical signs (corneal staining, gland expressibility, telangiectasia, vascularity, lash loss); however, there was a high correlation (r = 0.77; p < 0.05) between meibography score with percentage dropout.
CONCLUSIONS: Observers graded from -1 to +1 grade units between and within themselves for a four-point scale, 95% of the time. Although the interobserver and intraobserver repeatability of the K4 and K5M were very similar, a high rate of disagreement in percentage dropout between K4 and K5M images suggests that the two instruments cannot be interchanged. Meibomian gland dropout scores did not correlate significantly with clinical signs. Using a finer scale may be beneficial for detecting change.

Entities:  

Mesh:

Year:  2014        PMID: 24830370     DOI: 10.1097/OPX.0000000000000279

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  13 in total

Review 1.  Dysfunctional tear syndrome: dry eye disease and associated tear film disorders - new strategies for diagnosis and treatment.

Authors:  Mark S Milner; Kenneth A Beckman; Jodi I Luchs; Quentin B Allen; Richard M Awdeh; John Berdahl; Thomas S Boland; Carlos Buznego; Joseph P Gira; Damien F Goldberg; David Goldman; Raj K Goyal; Mitchell A Jackson; James Katz; Terry Kim; Parag A Majmudar; Ranjan P Malhotra; Marguerite B McDonald; Rajesh K Rajpal; Tal Raviv; Sheri Rowen; Neda Shamie; Jonathan D Solomon; Karl Stonecipher; Shachar Tauber; William Trattler; Keith A Walter; George O Waring; Robert J Weinstock; William F Wiley; Elizabeth Yeu
Journal:  Curr Opin Ophthalmol       Date:  2017-01       Impact factor: 3.761

2.  Tear eicosanoids in healthy people and ocular surface disease.

Authors:  Yohannes Abere Ambaw; Cecilia Chao; Shanshan Ji; Manfred Raida; Federico Torta; Markus R Wenk; Louis Tong
Journal:  Sci Rep       Date:  2018-07-26       Impact factor: 4.379

3.  A novel noninvasive ocular surface analyzer for the assessment of dry eye with Meibomian gland dysfunction.

Authors:  Yuanyuan Qi; Chen Zhang; Shaozhen Zhao; Yue Huang; Ruibo Yang
Journal:  Exp Ther Med       Date:  2017-04-19       Impact factor: 2.447

4.  Analysis of Age, Gender, and Refractive Error-Related Changes of the Anterior Corneal Surface Parameters Using Oculus Keratograph Topography.

Authors:  Nasrin Moghadas Sharif; Negareh Yazdani; Leila Shahkarami; Hadi Ostadi Moghaddam; Asieh Ehsaei
Journal:  J Curr Ophthalmol       Date:  2020-07-04

Review 5.  A Review of Imaging Biomarkers of the Ocular Surface.

Authors:  William W Binotti; Betul Bayraktutar; M Cuneyt Ozmen; Stephanie M Cox; Pedram Hamrah
Journal:  Eye Contact Lens       Date:  2020-03       Impact factor: 3.152

6.  Deficiency in Acyl-CoA:Wax Alcohol Acyltransferase 2 causes evaporative dry eye disease by abolishing biosynthesis of wax esters.

Authors:  Made Airanthi K Widjaja-Adhi; Josie A Silvaroli; Sylwia Chelstowska; Thomas Trischman; Ilya Bederman; Rony Sayegh; Marcin Golczak
Journal:  FASEB J       Date:  2020-08-26       Impact factor: 5.834

7.  Evaluation of Dry Eye and Meibomian Gland Dysfunction in Teenagers with Myopia through Noninvasive Keratograph.

Authors:  Xiu Wang; Xiaoxiao Lu; Jun Yang; Ruihua Wei; Liyuan Yang; Shaozhen Zhao; Xilian Wang
Journal:  J Ophthalmol       Date:  2016-01-06       Impact factor: 1.909

8.  The relief of dry eye signs and symptoms using a combination of lubricants, lid hygiene and ocular nutraceuticals.

Authors:  William Ngo; Sruthi Srinivasan; Diane Houtman; Lyndon Jones
Journal:  J Optom       Date:  2016-07-12

9.  An Eyelid Warming Device for the Management of Meibomian Gland Dysfunction.

Authors:  William Ngo; Sruthi Srinivasan; Lyndon Jones
Journal:  J Optom       Date:  2018-10-16

10.  Evaluation of ocular surface impairment in meibomian gland dysfunction of varying severity using a comprehensive grading scale.

Authors:  Jiayu Fu; Yilin Chou; Ran Hao; Xiaodan Jiang; Yushi Liu; Xuemin Li
Journal:  Medicine (Baltimore)       Date:  2019-08       Impact factor: 1.817

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.