Literature DB >> 24828214

Effects of hearing aid settings for electric-acoustic stimulation.

Margaret T Dillon1, Emily Buss1, Harold C Pillsbury1, Oliver F Adunka1, Craig A Buchman1, Marcia C Adunka2.   

Abstract

BACKGROUND: Cochlear implant (CI) recipients with postoperative hearing preservation may utilize an ipsilateral bimodal listening condition known as electric-acoustic stimulation (EAS). Studies on EAS have reported significant improvements in speech perception abilities over CI-alone listening conditions. Adjustments to the hearing aid (HA) settings to match prescription targets routinely used in the programming of conventional amplification may provide additional gains in speech perception abilities.
PURPOSE: Investigate the difference in users' speech perception scores when listening with the recommended HA settings for EAS patients versus HA settings adjusted to match National Acoustic Laboratories' nonlinear fitting procedure version 1 (NAL-NL1) targets. RESEARCH
DESIGN: Prospective analysis of the influence of HA settings. STUDY SAMPLE: Nine EAS recipients with greater than 12 mo of listening experience with the DUET speech processor. INTERVENTION: Subjects were tested in the EAS listening condition with two different HA setting configurations. Speech perception materials included consonant-nucleus-consonant (CNC) words in quiet, AzBio sentences in 10-talker speech babble at a signal-to-noise ratio (SNR) of +10, and the Bamford-Kowal-Bench sentences in noise (BKB-SIN) test. DATA COLLECTION AND ANALYSIS: The speech perception performance on each test measure was compared between the two HA configurations.
RESULTS: Subjects experienced a significant improvement in speech perception abilities with the HA settings adjusted to match NAL-NL1 targets over the recommended HA settings.
CONCLUSIONS: EAS subjects have been shown to experience improvements in speech perception abilities when listening to ipsilateral combined stimulation. This population's abilities may be underestimated with current HA settings. Tailoring the HA output to the patient's individual hearing loss offers improved outcomes on speech perception measures. American Academy of Audiology.

Entities:  

Mesh:

Year:  2014        PMID: 24828214     DOI: 10.3766/jaaa.25.2.2

Source DB:  PubMed          Journal:  J Am Acad Audiol        ISSN: 1050-0545            Impact factor:   1.664


  5 in total

1.  Combined Electric and Acoustic Stimulation With Hearing Preservation: Effect of Cochlear Implant Low-Frequency Cutoff on Speech Understanding and Perceived Listening Difficulty.

Authors:  René H Gifford; Timothy J Davis; Linsey W Sunderhaus; Christine Menapace; Barbara Buck; Jillian Crosson; Lori O'Neill; Anne Beiter; Phil Segel
Journal:  Ear Hear       Date:  2017 Sep/Oct       Impact factor: 3.570

2.  Effect of Place-Based Versus Default Mapping Procedures on Masked Speech Recognition: Simulations of Cochlear Implant Alone and Electric-Acoustic Stimulation.

Authors:  Margaret T Dillon; Brendan P O'Connell; Michael W Canfarotta; Emily Buss; Joseph Hopfinger
Journal:  Am J Audiol       Date:  2022-04-08       Impact factor: 1.636

3.  Effectiveness of Place-based Mapping in Electric-Acoustic Stimulation Devices.

Authors:  Margaret T Dillon; Michael W Canfarotta; Emily Buss; Joseph Hopfinger; Brendan P O'Connell
Journal:  Otol Neurotol       Date:  2021-01       Impact factor: 2.311

4.  Contributions of electric and acoustic hearing to bimodal speech and music perception.

Authors:  Joseph D Crew; John J Galvin; David M Landsberger; Qian-Jie Fu
Journal:  PLoS One       Date:  2015-03-19       Impact factor: 3.240

5.  Mandarin speech perception in combined electric and acoustic stimulation.

Authors:  Yongxin Li; Guoping Zhang; John J Galvin; Qian-Jie Fu
Journal:  PLoS One       Date:  2014-11-11       Impact factor: 3.240