Role of sinus in prosthetic venous valve.

W-H Tien1, H Y Chen2, Z C Berwick3, J Krieger4, S Chambers4, D Dabiri1, G S Kassab5.   

Abstract

BACKGROUND: The majority of bioprosthetic venous valves do not have a sinus pocket and, in practice, they are often placed in non-sinus segments of the veins. The aim of this study is to investigate the effect of the sinus pocket on the flow dynamics in a prosthetic valve.
METHODS: A bench top in vitro experiment was set up at physiological flow conditions to simulate the flow inside a venous system. Bicuspid bioprosthetic valves with different leaflet lengths (5 and 10 mm) were tested in tubes with and without a sinus pocket and the flows around the valve were visualized by particle image velocimetry (PIV). Velocity data measurements were made and the vorticity was calculated in the with- and without-sinus set-ups.
RESULTS: PIV measurements showed that vortex structure was maintained by the sinus. For the 10-mm leaflet length design with sinus, the jet width at the exit of the valve was 59% of that without sinus. For the 5-mm design with sinus, the jet width was 73% of the valve without sinus. Flow from the sinus region was entrained into the main jet observed near the exit of the sinus and altered the flow at the near wall region.
CONCLUSIONS: The sinus pocket alters the flow around the valve and functions as flow regulator to smooth the flow pattern around the valve. The vortical structure inside the sinus is maintained at the valve leaflet tip during the valve cycle. For the prosthetic valve designated to be placed without a sinus, a shorter leaflet length is preferable and performs more closely to the valve with sinus.
Copyright © 2014 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

Keywords:  Bioprosthetic valve; Flow dynamics; Particle image velocimetry; Pulse duplicator; Vortex

Mesh:

Year:  2014        PMID: 24820915     DOI: 10.1016/j.ejvs.2014.03.041

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  2 in total

1.  In silico analyses of blood flow and oxygen transport in human micro-veins and valves.

Authors:  Navaneeth Krishna Rajeeva Pandian; Abhishek Jain
Journal:  Clin Hemorheol Microcirc       Date:  2022       Impact factor: 2.411

2.  The effect of pathologic venous valve on neighboring valves: fluid-structure interactions modeling.

Authors:  Elina Soifer; Dar Weiss; Gil Marom; Shmuel Einav
Journal:  Med Biol Eng Comput       Date:  2016-09-23       Impact factor: 2.602

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.