Literature DB >> 24817563

Comparison of novel and familiar commercial kits for detection of C-reactive protein levels.

Ihsan Hakkı Ciftci1, Mehmet Koroglu, Engin Karakece.   

Abstract

In this study, randomized patient sera were used to simultaneously evaluate an automated C-reactive protein (CRP) assay and a commercial semi-automated microCRP assay with respect to correlation, linearity, and accuracy. Patient specimens were analyzed; two independent assay runs were performed on i-CHROMA (Boditech Med Inc., Korea) and IMMAGE 800 (Beckman Coulter Inc., USA) analyzers to estimate the between- and within-run precision. All systems were calibrated, and quality-control materials were analyzed according to the manufacturer's instructions. The results using the control materials were within the respective manufacturers' specified limits. The comparison studies were designed using the CLSI EP9-2A guidelines. The mean serum CRP concentrations were 123.2 ± 123.5 mg/L (95 % confidence of interval (CI) 97.9-148.3) using the CRP assay and 130.1 ± 109.3 mg/L (95 % CI 107.9-152.4) using the microCRP assay. The variance values were σ = 15,252.6 and 11,935.8 for the CRP and microCRP assays, respectively. The concordance correlation coefficient value was calculated as 0.8314 (95 % CI 0.7594-0.8833). There was a significant correlation between the CRP and microCRP assays: r = 0.8392 and 95 % CI 0.7675-0.8902 (p < 0.0001). The CRP and microCRP detection methods were well correlated. The i-CHROMA has many advantages over the IMMAGE 800 with respect to space required, analysis time, and system setup/application costs in a laboratory. It may be an attractive instrument for small and intermediate medical centers.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24817563     DOI: 10.1007/s11274-014-1653-9

Source DB:  PubMed          Journal:  World J Microbiol Biotechnol        ISSN: 0959-3993            Impact factor:   3.312


  8 in total

Review 1.  C-reactive protein: a critical update.

Authors:  Mark B Pepys; Gideon M Hirschfield
Journal:  J Clin Invest       Date:  2003-06       Impact factor: 14.808

Review 2.  Fluoroimmunoassays and immunofluorometric assays.

Authors:  I Hemmilä
Journal:  Clin Chem       Date:  1985-03       Impact factor: 8.327

3.  One-step homogeneous C-reactive protein assay for saliva.

Authors:  Chamindie Punyadeera; Goce Dimeski; Karam Kostner; Peter Beyerlein; Justin Cooper-White
Journal:  J Immunol Methods       Date:  2011-07-27       Impact factor: 2.303

4.  Comparison and evaluation of two C-reactive protein assays based on particle-enhanced immunoturbidimetry.

Authors:  Ali Riza Sişman; Tuncay Küme; Gültekin Taş; Pinar Akan; Pinar Tuncel
Journal:  J Clin Lab Anal       Date:  2007       Impact factor: 2.352

5.  Comparative study of two automated high-sensitivity C-reactive protein methods in a large population.

Authors:  Porntip H Lolekha; Anchalee Chittamma; William L Roberts; Piyamitr Sritara; Sayan Cheepudomwit; Paibul Suriyawongpaisal
Journal:  Clin Biochem       Date:  2005-01       Impact factor: 3.281

6.  C reactive protein rapid assay techniques for monitoring resolution of infection in immunosuppressed patients.

Authors:  R I Harris; P C Stone; A G Hudson; J Stuart
Journal:  J Clin Pathol       Date:  1984-07       Impact factor: 3.411

7.  Magnetic bead based assay for C-reactive protein using quantum-dot fluorescence labeling and immunoaffinity separation.

Authors:  Xiaoshan Zhu; Dayue Duan; Nelson G Publicover
Journal:  Analyst       Date:  2009-12-10       Impact factor: 4.616

8.  Serum C-reactive protein and neopterin concentrations in patients with viral or bacterial infection.

Authors:  A C Shaw
Journal:  J Clin Pathol       Date:  1991-07       Impact factor: 3.411

  8 in total
  2 in total

1.  Comparative analysis of high CRP-levels in human blood using point-of-care and laboratory-based methods.

Authors:  Christiane Eckschlager; Karin Schwenoha; Caroline Roth; Barbara Bogner; Gertie Janneke Oostingh
Journal:  Pract Lab Med       Date:  2019-09-19

2.  Performance evaluation of routine blood and C-reactive protein analysis using Mindray BC-7500 CRP auto hematology analyzer.

Authors:  Zhen Lin; Qiu Lin; Pingli Yu; Zhixin Chen; Haifeng Lin; Bin Zhu; Meihua Wang; Yingping Cao
Journal:  Ann Transl Med       Date:  2022-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.