Literature DB >> 24817449

Why vascular access trials on flow surveillance failed.

Nikolai Krivitski1.   

Abstract

INTRODUCTION: Since the introduction of access flow surveillance technology for routine patient screening in 1995, more than 30 clinical trials have been presented in peer reviewed journals. Despite overall positive outcomes, some trials, including randomized control trials (RCTs), failed to produce positive outcomes for access surveillance. The purpose of this study is to analyze published data related to the main component of access surveillance--adequate increase of access flow after percutaneous transluminal angioplasty (PTA).
RESULTS: A total of nine studies for arteriovenous grafts (AVGs) that include 350 accesses and nine studies for arteriovenous fistula (AVF) that included 503 accesses were considered for analysis from 14 publications. Practically, all reference data find high sensitivity (>90%) of access flow measurement to predict 50% stenosis. Mean access flow increase after PTA in AVGs was 319 ml/min (from 238 to 524 ml/min). Mean access flow increase in AVFs was 331 ml/min (from 195 to 402 ml/min). Relative flow increase in AVFs was 1.6 times greater than in AVGs. The authors of failed RCT for AVGs either did not select patients for PTA based on KDOQI guidelines and did not provide/analyze PTA flow results data, or reveal data that obviously show failure of PTA to adequately improve access flow.
SUMMARY: Access flow surveillance successfully identifies patients with hemodynamically significant stenosis. PTA performed on AVFs produce better hemodynamic results than in AVGs. Inadequate flow increases during PTA and not following KDOQI guidelines are major contributing factors for failed AVG randomized tails. Radiologists should use objective means for flow evaluation during PTA.

Entities:  

Mesh:

Year:  2014        PMID: 24817449     DOI: 10.5301/jva.5000256

Source DB:  PubMed          Journal:  J Vasc Access        ISSN: 1129-7298            Impact factor:   2.283


  2 in total

1.  Flexible, Skin Coupled Microphone Array for Point of Care Vascular Access Monitoring.

Authors:  Binit Panda; Soumyajit Mandal; Steve J A Majerus
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2019-10-18       Impact factor: 3.833

2.  High hemodialysis vascular access flow and impaired right ventricular function in chronic hemodialysis patients.

Authors:  S Yilmaz; M Yetim; B K Yilmaz; T Dogan; E Aksoy; N Yuksel; I Dogan
Journal:  Indian J Nephrol       Date:  2016-09
  2 in total