Literature DB >> 24815184

Regional strain determination and myocardial infarction detection by three-dimensional echocardiography with varied temporal resolution.

Meihua Zhu1, Cole Streiff, Jill Panosian, Zhijun Zhang, Xubo Song, David J Sahn, Muhammad Ashraf.   

Abstract

BACKGROUND: Three-dimensional echocardiography (3DE) is a promising method for strain determination; however, there are temporal resolution concerns. This study aims to evaluate the feasibility and accuracy of 3DE on longitudinal and circumferential strain (LS, CS) determination and infarction detection under variable frame rates (FR) and "heart rates" (stroke rates [SR]) conditions.
METHODS: Latex balloons were sewn into the left ventricle (LV) of 20 freshly harvested pig hearts which were then passively driven by a pulsatile pump apparatus at stroke volumes (SV) 30-70 mL. The hearts were pumped at 2 normal limits of human heart rate. Full-volume data were acquired before and after a simulated myocardial infarction (MI) at the 2 most commonly used FRs. LS and CS values were evaluated against sonomicrometry.
RESULTS: Longitudinal strain and CS derived from high FR acquisitions showed statistically superior correlations with sonomicrometry data (LS: R(2) = 0.85, CS: R(2) = 0.84) than strain values from low FR (LS: R(2) = 0.78, CS: R(2) = 0.76) (all P < 0.01). After MI induction, LS and CS at different FRs were significantly decreased while maintaining excellent correlations with sonomicrometry data (all P < 0.001). There is no statistical difference of strain values between different SR acquisitions.
CONCLUSION: Three-dimensional wall-motion tracking has the ability to accurately determine regional myocardial deformation and detect MI. Different heart rates within a physiologically relevant range have no effect on 3D strain accuracy. Strain values calculated from higher frame rate acquisitions were found to have a slightly better accuracy.
© 2014, Wiley Periodicals, Inc.

Entities:  

Keywords:  myocardial infarction; strain; temporal resolution; wall motion

Mesh:

Year:  2014        PMID: 24815184     DOI: 10.1111/echo.12632

Source DB:  PubMed          Journal:  Echocardiography        ISSN: 0742-2822            Impact factor:   1.724


  4 in total

1.  Intraobserver and interobserver reproducibility for radial, circumferential and longitudinal strain echocardiography.

Authors:  R Leischik; B Dworrak; K Hensel
Journal:  Open Cardiovasc Med J       Date:  2014-10-17

2.  Non-Invasive Evaluation of Heart Function with Four-Dimensional Echocardiography.

Authors:  Ran Chen; Meihua Zhu; David J Sahn; Muhammad Ashraf
Journal:  PLoS One       Date:  2016-05-04       Impact factor: 3.240

3.  Evaluation of left and right ventricular myocardial function after lung resection using speckle tracking echocardiography.

Authors:  Zhenhua Wang; Jianjun Yuan; Wen Chu; Yuhong Kou; Xijun Zhang
Journal:  Medicine (Baltimore)       Date:  2016-08       Impact factor: 1.889

4.  Cardiac Mechanics in Isolated Bicuspid Aortic Valve Disease With Normal Ejection Fraction: A Study of Various Valvular Lesion Types.

Authors:  Xijun Zhang; Meihua Zhu; Tao He; Jianjun Yuan; Haohui Zhu; Dennis E Morrisroe; Muhammad Ashraf; David J Sahn
Journal:  Medicine (Baltimore)       Date:  2015-11       Impact factor: 1.817

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.