Simulated cystic renal lesions: quantitative X-ray phase-contrast CT--an in vitro phantom study.

Alexander A Fingerle1, Marian Willner, Julia Herzen, Daniela Münzel, Dieter Hahn, Ernst J Rummeny, Peter B Noël, Franz Pfeiffer.   

Abstract

PURPOSE: To determine if grating-based x-ray phase-contrast computed tomography (CT) can allow differentiation of simulated simple, protein-rich, hemorrhagic, and enhancing cystic renal lesions in an in vitro phantom.
MATERIALS AND METHODS: An in vitro phantom specifically designed to simulate simple, protein-rich, hemorrhagic, and enhancing renal cysts was scanned with an experimental grating-based phase-contrast CT setup consisting of a Talbot-Lau interferometer with a rotating anode x-ray tube and a single photon counting detector. Various combinations of serum and saline (100% and 0% to 0% and 100%), blood and saline, blood and serum (100% and 0% to 6.25% and 93.75% for both), and an iodinated contrast agent and saline (7.6-1.6 mg per milliliter of saline) were used to reproduce the chemical composition of the different types of cysts. A thickened solution of an iodinated contrast agent calibrated with a clinical multidetector CT scanner served as contrast agent-enhanced renal parenchyma (195 HU at 80 kVp, 400 mAs and 98 HU at 140 kVp, 200 mAs). Standard attenuation- and phase-contrast images were reconstructed from the raw projection data. Quantitative values for attenuation and phase contrast and image noise were determined. Contrast-to-noise ratios were calculated. Simulated lesions were assessed for visual differentiability by means of pairwise comparison of the attenuation- and phase-contrast images and both images simultaneously.
RESULTS: Attenuation-contrast imaging showed large differences in Hounsfield units with increasing concentrations of iodine (118.9 HU for 1.6 mg/mL vs 331.4 HU for 7.6 mg/mL). Values for phase-contrast imaging were substantially distinguishable for saline, serum, and blood (7.9, 23.7, and 52.8 HU, respectively). Both imaging modalities combined allowed differentiation of all four types of simulated cysts (100% saline, 100% serum, 100% blood, and 1.6-7.6 mg of iodine per milliliter of saline) with one imaging acquisition.
CONCLUSION: Grating-based phase-contrast CT allows differentiation of simulated simple, protein-rich, hemorrhagic, and enhancing renal cysts in an in vitro phantom through simultaneous assessment of their distinct attenuation- and phase-contrast signal.

Entities:  

Mesh:

Year:  2014        PMID: 24814175     DOI: 10.1148/radiol.14130876

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  6 in total

1.  Multi-Mounted X-Ray Computed Tomography.

Authors:  Jian Fu; Zhenzhong Liu; Jingzheng Wang
Journal:  PLoS One       Date:  2016-04-13       Impact factor: 3.240

2.  High-resolution 3D visualization of ductular proliferation of bile duct ligation-induced liver fibrosis in rats using x-ray phase contrast computed tomography.

Authors:  Lili Qin; Xinyan Zhao; Jianbo Jian; Yuqing Zhao; Mengyu Sun; Chunhong Hu
Journal:  Sci Rep       Date:  2017-06-26       Impact factor: 4.379

3.  X-ray Phase Contrast osteo-articular imaging: a pilot study on cadaveric human hands.

Authors:  Hélène Rougé-Labriet; Sebastien Berujon; Hervé Mathieu; Sylvain Bohic; Barbara Fayard; Jean-Noel Ravey; Yohann Robert; Philippe Gaudin; Emmanuel Brun
Journal:  Sci Rep       Date:  2020-02-05       Impact factor: 4.379

4.  Experimental Realisation of High-sensitivity Laboratory X-ray Grating-based Phase-contrast Computed Tomography.

Authors:  Lorenz Birnbacher; Marian Willner; Astrid Velroyen; Mathias Marschner; Alexander Hipp; Jan Meiser; Frieder Koch; Tobias Schröter; Danays Kunka; Jürgen Mohr; Franz Pfeiffer; Julia Herzen
Journal:  Sci Rep       Date:  2016-04-04       Impact factor: 4.379

5.  High resolution laboratory grating-based X-ray phase-contrast CT.

Authors:  Manuel Viermetz; Lorenz Birnbacher; Marian Willner; Klaus Achterhold; Franz Pfeiffer; Julia Herzen
Journal:  Sci Rep       Date:  2018-10-26       Impact factor: 4.379

Review 6.  Quantitative X-ray phase contrast computed tomography with grating interferometry : Biomedical applications of quantitative X-ray grating-based phase contrast computed tomography.

Authors:  Lorenz Birnbacher; Eva-Maria Braig; Daniela Pfeiffer; Franz Pfeiffer; Julia Herzen
Journal:  Eur J Nucl Med Mol Imaging       Date:  2021-04-13       Impact factor: 9.236

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.