BACKGROUND AND
OBJECTIVE: To evaluate the usefulness of a 20-gauge cannula to maintain a self-sealing sclerotomy wound after 23-gauge phacofragmentation. PATIENTS AND METHODS: This retrospective study compared the suture rates after 23-gauge phacofragmentation when the 23-gauge cannula was temporarily replaced with a 20-gauge valved metal cannula versus when the 23-gauge fragmatome was inserted at the sclerotomy site without a cannula.
RESULTS: Whereas a sclerotomy was sutured in all 31 eyes in the without-cannula group, only one eye of 14 in the cannula group required a sclerotomy suture (P < .0001). However, there was one case of fragmatome tip fracture during fragmentation in the cannula group.
CONCLUSION: A self-sealing sclerotomy wound can be maintained without suturing by using a 20-gauge metal cannula, but fragmatome tip fracture can occur during fragmentation. Copyright 2014, SLACK Incorporated.

Entities:  

Mesh:

Year:  2014        PMID: 24806268     DOI: 10.3928/23258160-20140429-01

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers Imaging Retina        ISSN: 2325-8160            Impact factor:   1.300


  2 in total

1.  Computational Modeling of Ophthalmic Procedures: Computational Modeling of Ophthalmic Procedures.

Authors:  William J Foster; Brian W Berg; Steven N Luminais; Amir Hadayer; Shlomit Schaal
Journal:  Am J Ophthalmol       Date:  2022-03-28       Impact factor: 5.488

2.  Design Parameters for a Small-Gauge Fragmatome.

Authors:  William J Foster; Jijo Jizhou Wang
Journal:  Transl Vis Sci Technol       Date:  2019-08-07       Impact factor: 3.283

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.