Literature DB >> 24806140

The sevoflurane washout profile of seven recent anesthesia workstations for malignant hyperthermia-susceptible adults and infants: a bench test study.

Nicolas Cottron1, Claire Larcher, Agnès Sommet, Rose Fesseau, Xavier Alacoque, Vincent Minville, Olivier Fourcade, Delphine Kern.   

Abstract

BACKGROUND: Preoperative flushing of an anesthesia workstation is an alternative for preparation of the anesthesia workstation before use in malignant hyperthermia-susceptible patients (MHS). We studied in vitro, using a test lung, the washout profile of sevoflurane in 7 recent workstations during adult and, for the first time, pediatric ventilation patterns.
METHODS: Anesthesia workstations were first primed with 3% sevoflurane for 2 hours and then prepared according to the recommendations of the Malignant Hyperthermia Association of the United States. The flush was done with maximal fresh gas flow (FGF) with a minute ventilation equal to 600 mL × 15, to reach a sevoflurane concentration of <5 parts per million. After flush, 2 clinical situations were simulated in vitro to test the efficiency of preparation: decrease of FGF from max to 10 L/min, or decrease of minute ventilation to 50 mL × 30, to simulate the ventilation of an MHS infant.
RESULTS: We report washout delays for MHS patients for previously studied workstations (Primus®, Avance®, and Zeus®) and more interestingly, for machines not previously tested (Felix®, Flow-I®, Perseus®, and Leon®). An increase of sevoflurane concentration was observed when decreasing FGF (except for flow-I® and Leon®) and during simulation of MHS infant ventilation (except for Felix®).
CONCLUSIONS: This descriptive study strongly suggests that washout profiles may differ for each anesthesia workstation. We advise the use of maximal FGF during preparation and anesthesia. Required flushing times are longer when preparing an anesthesia workstation before providing anesthesia for MHS infants.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24806140     DOI: 10.1213/ANE.0000000000000208

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  2 in total

1.  Seizure-like movements caused by residual sevoflurane inside the anesthesia machine: A case report.

Authors:  Jianqiao Zheng; Li Du; Lu Zhang
Journal:  Medicine (Baltimore)       Date:  2021-01-29       Impact factor: 1.817

2.  Preparation of Dräger Atlan A350 and General Electric Healthcare Carestation 650 anesthesia workstations for malignant hyperthermia susceptible patients.

Authors:  Sebastian Heiderich; Christian Thoben; Nils Dennhardt; Terence Krauß; Robert Sümpelmann; Stefan Zimmermann; Michael Reitz; Henrik Rüffert
Journal:  BMC Anesthesiol       Date:  2021-12-13       Impact factor: 2.217

  2 in total