Literature DB >> 24802626

Randomized controlled trial of perioperative ICD management: magnet application versus reprogramming.

Janet Gifford1, Karen Larimer, Celia Thomas, Patricia May, Stephen Stanhope, Apoor Gami.   

Abstract

BACKGROUND: There are insufficient data to guide perioperative implantable cardioverter-defibrillator (ICD) management for patients undergoing surgical procedures using electrocautery.
METHODS: We conducted a multicenter randomized controlled trial of patients with ICDs undergoing surgery with monopolar electrocautery. Subjects were randomized to an "Off" group (ICD therapy programmed off, then postoperatively programmed on) or a "Magnet" group (ICD therapy suspended with a magnet and no immediate postoperative ICD interrogation). Also, a registry was maintained of ICD patients with procedures within 6 inches of the ICD (all programmed off). The primary endpoint was ICD off time with secondary endpoints being caregiver handoffs and incidence of electromagnetic interference (EMI).
RESULTS: All patients (n = 80) had pectoral ICDs. Subject demographics were well matched in each group, and duration of electrocautery was similar (80 minutes vs 64 minutes, P = 0.58). The mean "excess" ICD off time (ICD off time - electrocautery time) was significantly higher in the Off group than the Magnet group (115 minutes vs 28 minutes, P < 0.001). Mean number of caregiver handoffs were higher in the Off group (6.6 vs 5.5, P < 0.001). There was no EMI in any lower abdominal or lower extremity procedures. Neither group had arrhythmic events or device reset.
CONCLUSION: A magnet protocol simplifies perioperative ICD management for procedures using electrocautery more than 6 inches from the ICD. This protocol results in significantly shorter time with ICD therapy off, fewer provider handoffs, no risk of inadvertently discharging patients home with ICD therapies off, and no device reset. ©2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  clinical trials; defibrillation - ICD; surgery

Mesh:

Year:  2014        PMID: 24802626     DOI: 10.1111/pace.12417

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  6 in total

1.  A randomized in vitro evaluation of transient and permanent cardiac implantable electronic device malfunctions following direct exposure up to 10 Gy.

Authors:  Maria Daniela Falco; Domenico Genovesi; Luciana Caravatta; Clelia Di Carlo; Ekaterina Bliakharskaia; Marianna Appignani; Massimiliano Faustino; Nanda Furia; Enrico Di Girolamo
Journal:  Strahlenther Onkol       Date:  2020-06-19       Impact factor: 3.621

Review 2.  Management of Patients With Cardiovascular Implantable Electronic Devices in Dental, Oral, and Maxillofacial Surgery.

Authors:  James Tom
Journal:  Anesth Prog       Date:  2016

3.  Electromagnetic interference in implantable cardioverter defibrillators: present but rare.

Authors:  Gesa von Olshausen; Ina-Christine Rondak; Carsten Lennerz; Verena Semmler; Christian Grebmer; Tilko Reents; Sonia Ammar-Busch; Alessandra Buiatti; Felix Bourier; Isabel Deisenhofer; Christof Kolb
Journal:  Clin Res Cardiol       Date:  2016-01-29       Impact factor: 5.460

4.  Position Statement on the Management of Cardiac Electrophysiology and Cardiac Implantable Electronic Devices in Australia During the COVID-19 Pandemic: A Living Document.

Authors:  Saurabh Kumar; Haris Haqqani; Gareth Wynn; Rajeev K Pathak; Jonathan Lipton; Rajiv Mahajan; Prashanthan Sanders; Stewart Healey; Bradley Wilsmore; Justin A Mariani; Stuart P Thomas; Rukshen Weerasooriya; Andrew McGavigan; Paul A Gould; Paul Weatherley; Natasha Saad; Mitchell Cowan; Samual Turnbull; Ivana Trivic; Michael Wong; Ivaylo Tonchev; Joseph B Morton; Jonathan R Skinner; Andreas Pflaumer; Mark McGuire; Peter Kistler; Jonathan M Kalman
Journal:  Heart Lung Circ       Date:  2020-05-11       Impact factor: 2.975

Review 5.  Perioperative Management of Patients with Cardiac Implantable Electronic Devices and Utility of Magnet Application.

Authors:  Tardu Özkartal; Andrea Demarchi; Maria Luce Caputo; Enrico Baldi; Giulio Conte; Angelo Auricchio
Journal:  J Clin Med       Date:  2022-01-28       Impact factor: 4.241

Review 6.  State-of-the-art consensus on non-transvenous implantable cardioverter-defibrillator therapy.

Authors:  Christoph Schukro; David Santer; Günther Prenner; Markus Stühlinger; Martin Martinek; Alexander Teubl; Deddo Moertl; Stefan Schwarz; Michael Nürnberg; Lukas Fiedler; Robert Hatala; Cesar Khazen
Journal:  Clin Cardiol       Date:  2020-08-14       Impact factor: 2.882

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.