Evaluation of the Analytical Performance of the Olympus AU5421 Applied with Parameters Developed by the Toshiba 200FR for Measurement of Blood Ethanol Concentration.

Bo Hyun Kim1, Ae Ja Park1, Jae Saeng Youn2, Jae Hoon Bae2.   

Abstract

BACKGROUND: An accurate determination of blood ethanol concentrations is important. To minimize ethanol degradation in blood samples, sodium fluoride (NaF) collection tubes have been recommended for use. In this study, we attempted to utilize the Olympus AU5421 chemistry analyzer for ethanol analysis based on the parameters established for the Toshiba 200FR. We also evaluated the effect of NaF collection tubes on ethanol concentrations.
METHODS: The precision, linearity, accuracy, and carry-over rate of the AU5421 analyzer were evaluated. The results of analysis using the AU5421 and Abbott AxSYM analyzers were also compared. The effects of NaF collection tubes on ethanol concentrations in stored samples were measured.
RESULTS: The AU5421 showed a good precision, linearity, accuracy, and carry-over rate. The ethanol concentrations were well correlated with the results obtained using the AxSYM. There was no statistically significant difference in blood ethanol concentrations between the samples collected in tubes with NaF and those collected in tubes without NaF.
CONCLUSIONS: Since the AU5421 showed excellent analytical performance, the AU5421 could be used as an alternative to AxSYM for the determination of blood ethanol concentrations. Our analysis also indicated that there is no need to use NaF collection tubes if blood ethanol concentrations are analyzed within 3 h after blood collection. We believe that the results obtained in this study will have important implications for the use of the AU5421 system to measure blood ethanol concentrations.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  Olympus AU5421; Toshiba 200FR; alcohol; ethanol

Mesh:

Substances:

Year:  2014        PMID: 24798799      PMCID: PMC6807238          DOI: 10.1002/jcla.21753

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  5 in total

1.  Blood alcohol testing: comparison of the performance obtained with the different methods used in the Belgian external quality assessment schemes.

Authors:  Nicole Devleeschouwer; Jean Claude Libeer; Frank K Martens; Hugo Neels; Marc Van Damme; Alain Verstraete; Marc Deveaux; Pierre E Wallemacq
Journal:  Clin Chem Lab Med       Date:  2004-01       Impact factor: 3.694

2.  Electrochemical detection of blood alcohol concentration using a disposable biosensor based on screen-printed electrode modified with Nafion and gold nanoparticles.

Authors:  Peng Luo; Guoming Xie; Yi Liu; Huajian Xu; Shixiong Deng; Fangzhou Song
Journal:  Clin Chem Lab Med       Date:  2008       Impact factor: 3.694

3.  [Use of the ADH/REA method (Abbott Tdx-REA) in forensic blood alcohol determination].

Authors:  D Mebs; K Schmidt; H Bratzke
Journal:  Blutalkohol       Date:  1996-03

4.  The validity of police assessment of driver intoxication in motor vehicle crashes leading to hospitalization.

Authors:  D C Grossman; B A Mueller; T Kenaston; P Salzberg; W Cooper; G J Jurkovich
Journal:  Accid Anal Prev       Date:  1996-07

5.  [The accuracy of blood alcohol determination with head-space GC, ADH and REA ethanol assay for the AXSYM system--a comparison of the methods].

Authors:  A Alt; G Reinhardt
Journal:  Blutalkohol       Date:  1996-07
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.