| Literature DB >> 24793982 |
Lindsay B Baker1, Corey T Ungaro, Kelly A Barnes, Ryan P Nuccio, Adam J Reimel, John R Stofan.
Abstract
Abstract This study compared a field versus reference laboratory technique for extracting (syringe vs. centrifuge) and analyzing sweat [Na(+)] and [K(+)] (compact Horiba B-722 and B-731, HORIBA vs. ion chromatography, HPLC) collected with regional absorbent patches during exercise in a hot-humid environment. Sweat samples were collected from seven anatomical sites on 30 athletes during 1-h cycling in a heat chamber (33°C, 67% rh). Ten minutes into exercise, skin was cleaned/dried and two sweat patches were applied per anatomical site. After removal, one patch per site was centrifuged and sweat was analyzed with HORIBA in the heat chamber (CENTRIFUGE HORIBA) versus HPLC (CENTRIFUGE HPLC). Sweat from the second patch per site was extracted using a 5-mL syringe and analyzed with HORIBA in the heat chamber (SYRINGE HORIBA) versus HPLC (SYRINGE HPLC). CENTRIFUGE HORIBA, SYRINGE HPLC, and SYRINGE HORIBA were highly related to CENTRIFUGE HPLC ([Na(+)]: ICC = 0.96, 0.94, and 0.93, respectively; [K(+)]: ICC = 0.87, 0.92, and 0.84, respectively), while mean differences from CENTRIFUGE HPLC were small but usually significant ([Na(+)]: 4.7 ± 7.9 mEql/L, -2.5 ± 9.3 mEq/L, 4.0 ± 10.9 mEq/L (all P < 0.001), respectively; [K(+)]: 0.44 ± 0.52 mEq/L (P < 0.001), 0.01 ± 0.49 mEq/L (P = 0.77), 0.50 ± 0.48 mEq/L (P < 0.001), respectively). On the basis of typical error of the measurement results, sweat [Na(+)] and [K(+)] obtained with SYRINGE HORIBA falls within ±15.4 mEq/L and ±0.68 mEq/L, respectively, of CENTRIFUGE HPLC 95% of the time. The field (SYRINGE HORIBA) method of extracting and analyzing sweat from regional absorbent patches may be useful in obtaining sweat [Na(+)] when rapid estimates in a hot-humid field setting are needed.Entities:
Keywords: Compact electrolyte analyzers; Horiba; regional sweat collection; sweat patches
Year: 2014 PMID: 24793982 PMCID: PMC4098735 DOI: 10.14814/phy2.12007
Source DB: PubMed Journal: Physiol Rep ISSN: 2051-817X
Figure 1.Sample extraction and analysis scheme.
Intraday reliability of HPLC and HORIBA analyzers for sweat [Na+] and [K+].
| Sweat [Na+] | Sweat [K+] | |||||||
|---|---|---|---|---|---|---|---|---|
| CENTRIFUGE HPLC | CENTRIFUGE HORIBA | CENTRIFUGE HPLC | CENTRIFUGE HORIBA | |||||
| Replicate 1 vs. 2 | Replicate 2 vs. 3 | Replicate 1 vs. 2 | Replicate 2 vs. 3 | Replicate 1 vs. 2 | Replicate 2 vs. 3 | Replicate 1 vs. 2 | Replicate 2 vs. 3 | |
|
| 205 | 204 | 206 | 167 | 116 | 115 | 113 | 75 |
| Mean Difference ± SD (mEq/L) | −0.20 ± 1.73 | −0.09 ± 1.51 | 0.26 ± 1.41* | 0.10 ± 1.18 | 0.01 ± 0.18 | 0.00 ± 0.15 | −0.00 ± 0.14 | −0.01 ± 0.15 |
| 95% CI of Mean Difference (mEq/L) | −0.44 to 0.04 | −0.30 to 0.12 | 0.07 to 0.45 | −0.08 to 0.28 | −0.03 to 0.04 | −0.03 ± 0.03 | −0.03 to 0.02 | −0.04 to 0.03 |
| ICC | 0.998* | 0.999* | 0.999* | 0.999* | 0.988* | 0.992* | 0.996* | 0.996* |
| SEE (mEq/L) | 1.73 | 1.51 | 1.38 | 1.18 | 0.18 | 0.15 | 0.14 | 0.15 |
| TEM (mEq/L) | 1.22 | 1.07 | 1.00 | 0.83 | 0.13 | 0.11 | 0.10 | 0.11 |
| CV (%) | 1.97 | 1.67 | 1.35 | 1.17 | 3.27 | 2.81 | 2.35 | 2.21 |
To determine intraday reliability samples were measured in triplicate when possible (depending on sample volume; exact n for each replicate comparison is provided). CV, coefficient of variation; HPLC, ion chromatography using the Dionex ICS‐3000; HORIBA, Horiba B‐722 for sweat [Na+] and Horiba B‐731 for sweat [K+]; ICC, intraclass correlation coefficient (based on two‐way mixed ANOVA, absolute agreement, average measures); SEE, standard error of the estimate; TEM, typical error of measurement; *P < 0.001.
Interday reliability of HORIBA analyzer for sweat [Na+] and [K+].
| Sweat [Na+] | Sweat [K+] | |||
|---|---|---|---|---|
| Day 1 to day 2 | Day 2 to day 3 | Day 1 to day 2 | Day 2 to day 3 | |
| Mean difference ± SD (mEq/L) | −0.07 ± 3.34 | −0.08 ± 2.03 | −0.04 ± 0.39 | −0.09 ± 0.32 |
| 95% CI of mean difference (mEq/L) | −1.00 to 0.86 | −0.65 to 0.48 | −0.17 to 0.09 | −0.18 to 0.01 |
| ICC | 0.995* | 0.998* | 0.945* | 0.967* |
| SEE (mEq/L) | 2.76 | 1.94 | 0.36 | 0.32 |
| TEM (mEq/L) | 2.36 | 1.43 | 0.28 | 0.23 |
| CV (%) | 3.72 | 2.26 | 7.37 | 5.46 |
Interday reliability was measured across three repeat days in 52 and 46 samples for sweat [Na+] and [K+], respectively. CV, coefficient of variation; HORIBA, Horiba B‐722 for sweat [Na+] and Horiba B‐731 for sweat [K+]; ICC, intraclass correlation coefficient (based on two‐way mixed ANOVA, absolute agreement, average measures); SEE, standard error of the estimate; TEM, typical error of measurement.
Validity of Horiba analyzer: CENTRIFUGE HORIBA versus CENTRIFUGE HPLC for sweat [Na+] and [K+].
| Sweat [Na+] | Sweat [K+] | |
|---|---|---|
| Mean difference ± SD (mEq/L) | 4.71 ± 7.87* | 0.44 ± 0.52* |
| 95% CI of mean difference (mEq/L) | 3.64–5.78 | 0.34–0.54 |
| ICC | 0.96* | 0.87* |
| SEE (mEq/L) | 5.56 | 0.39 |
| TEM (mEq/L) | 5.56 | 0.37 |
| CV (%) | 9.37 | 8.36 |
n = 210 for sweat [Na+] and 116 for sweat [K+]. CV. coefficient of variation; CI, confidence interval; HPLC, ion chromatography using the Dionex ICS‐3000; HORIBA, Horiba B‐722 for sweat [Na+] and Horiba B‐731 for sweat [K+]; ICC, intraclass correlation coefficient (based on two‐way mixed ANOVA, absolute agreement, average measures); SD, standard deviation; SEE, standard error of the estimate; TEM, typical error of measurement; *P < 0.001.
Figure 2.Regression of CENTRIFUGE HORIBA (centrifuge sweat extraction, compact Horiba B‐722 analysis), SYRINGE HPLC (syringe sweat extraction, ion chromotography analysis), and SYRINGE HORIBA (combined field technique using syringe sweat extraction and compact Horiba B‐722 analysis) on CENTRIFUGE HPLC (reference laboratory technique using centrifuge sweat extraction and ion chromatography analysis) for sweat [Na+]. n = 210. Dashed line is the line of identity.
Figure 3.Regression of CENTRIFUGE HORIBA (centrifuge sweat extraction, compact Horiba B‐731 analysis), SYRINGE HPLC (syringe sweat extraction, HPLC analysis), and SYRINGE HORIBA (combined field technique using syringe sweat extraction and compact Horiba B‐731 analysis) on CENTRIFUGE HPLC (reference laboratory technique using centrifuge sweat extraction and ion chromatography analysis) for sweat [K+]. n = 116. Dashed line is the line of identity.
Validity of syringe sweat extraction: SYRINGE HPLC versus CENTRIFUGE HPLC for sweat [Na+] and [K+].
| Sweat [Na+] | Sweat [K+] | |
|---|---|---|
| Mean difference ± SD (mEq/L) | −2.52 ± 9.25* | 0.01 ± 0.49 |
| 95% CI of mean difference (mEq/L) | −3.78 to −1.26 | −0.08 to 0.10 |
| ICC | 0.94* | 0.92* |
| SEE (mEq/L) | 9.07 | 0.50 |
| TEM (mEq/L) | 5.51 | 0.35 |
| CV (%) | 11.20 | 9.22 |
n = 210 for sweat [Na+] and 116 for sweat [K+]. CV. coefficient of variation; CI, confidence interval; HPLC, ion chromatography using the Dionex ICS‐3000; HORIBA, Horiba B‐722 for sweat [Na+] and Horiba B‐731 for sweat [K+]; ICC, intraclass correlation coefficient (based on two‐way mixed ANOVA, absolute agreement, average measures); SD, standard deviation; SEE, standard error of the estimate; TEM, typical error of measurement; *P < 0.001.
Validity of field technique using syringe extraction of sweat and compact Horiba analysis: SYRINGE HORIBA versus CENTRIFUGE HPLC for sweat [Na+] and [K+].
| Sweat [Na+] | Sweat [K+] | |
|---|---|---|
| Mean difference ± SD (mEq/L) | 3.97 ± 10.87* | 0.50 ± 0.48* |
| 95% CI of mean difference (mEq/L) | 2.49–5.45 | 0.42–0.59 |
| ICC | 0.93* | 0.84* |
| SEE (mEq/L) | 8.68 | 0.44 |
| TEM (mEq/L) | 7.68 | 0.34 |
| CV (%) | 12.54 | 8.48 |
n = 210 for sweat [Na+] and 116 for sweat [K+]. CV. coefficient of variation; CI, confidence interval; HPLC, ion chromatography using the Dionex ICS‐3000; HORIBA, Horiba B‐722 for sweat [Na+] and Horiba B‐731 for sweat [K+]; ICC, intraclass correlation coefficient (based on two‐way mixed ANOVA, absolute agreement, average measures); SD, standard deviation; SEE, standard error of the estimate; TEM, typical error of measurement; *P < 0.001.