Evaluation of a third party enzymatic ammonia method for use on the Roche Cobas 6000 (c501) automated platform.

Isolde Seiden-Long1, Kareena Schnabl2, Wendy Skoropadyk2, Nola Lennon3, Arlayne McKeage3.   

Abstract

OBJECTIVE: Adaptation of the Randox Enzymatic Manual UV Ammonia method to be used on the Roche Cobas 6000 (c501) automated analyzer platform. DESIGN AND METHODS: The Randox ammonia reagent was evaluated for precision, linearity, accuracy and interference from hemolysis, icterus and lipemia on the Roche c501 analyzer. Comparison studies were conducted for the Randox reagent between Roche c501, Siemens Vista, Ortho Vitros 250, and Beckman DxC methods.
RESULTS: The Randox reagent demonstrates acceptable within-run (L1=65 μmol/L, CV 3.4% L2=168 μmol/L, CV 1.9%) and between-run precision (L1=29 μmol/L, CV 7.3% L2=102 μmol/L, CV 3.0%), Analytical Measurement Range (7-940 μmol/L), and accuracy. The method interference profile is superior for the Randox method (hemolysis index up to 600, icteric index up to 60, lipemic index up to 100) as compared to the Roche method (hemolysis index up to 200, icteric index up to 10, lipemic index up to 50). Comparison was very good between the Randox reagent and two other wet chemistry platforms.
CONCLUSIONS: The Randox Enzymatic Manual UV Ammonia reagent is an available alternative to the Roche Cobas c501 reagent. The method is more robust to endogenous interferences and less prone to instrument error flags, thus allowing the majority of clinical specimens to be reported without additional sample handling at our institution.
Copyright © 2014 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Ammonia; Application development; Automated analyzers; Enzymatic assays; Method evaluation

Mesh:

Substances:

Year:  2014        PMID: 24792455     DOI: 10.1016/j.clinbiochem.2014.04.022

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  1 in total

1.  Improved assay protocol for measurement of ammonia on the Roche Cobas 8000 automated platform.

Authors:  J Kaplon; J J de Groot; J P van Straalen; M Heckman; J C Fischer
Journal:  Pract Lab Med       Date:  2018-12-07
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.