Changes in morphology of the paced QRS complex related to pacemaker output.

L M van Gelder1, M I el Gamal, C H Tielen.   

Abstract

The influence of pacemaker output on the morphology of the paced QRS complex was studied from standard lead electrocardiograms in 69 patients with bipolar pacemakers. In 40 of the 69 patients (58%), there was a significant (P less than 0.001) change in electrical axis, from -75 degrees at the low output setting (2.7 V, 0.15 msec) to -67 degrees at the high output setting (8.1 V, 2.29 msec). In 30 patients, these changes were also associated with changes in the QRS morphology and in the T-wave. This phenomenon may be explained by additional stimulation from the proximal electrode at high output, thus altering the pattern of depolarization.

Entities:  

Mesh:

Year:  1989        PMID: 2477820     DOI: 10.1111/j.1540-8159.1989.tb01843.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Triple-site pacing in patients with biventricular device-incidence of the phenomenon and cardiac resynchronization benefit.

Authors:  Alan Bulava; Gerardo Ansalone; Renato Ricci; Paride Giannantoni; Carlo Pignalberi; Petr Heinc; Jan Lukl; Massimo Santini
Journal:  J Interv Card Electrophysiol       Date:  2004-02       Impact factor: 1.900

2.  Functional pace-mapping responses for identification of targets for catheter ablation of scar-mediated ventricular tachycardia.

Authors:  Roderick Tung; Nilesh Mathuria; Yoav Michowitz; Ricky Yu; Eric Buch; Jason Bradfield; Ravi Mandapati; Isaac Wiener; Noel Boyle; Kalyanam Shivkumar
Journal:  Circ Arrhythm Electrophysiol       Date:  2012-02-14

3.  Limitations and Challenges in Mapping Ventricular Tachycardia: New Technologies and Future Directions.

Authors:  Adam J Graham; Michele Orini; Pier D Lambiase
Journal:  Arrhythm Electrophysiol Rev       Date:  2017-08
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.