Use of the Nexfin™ device to detect acute arterial pressure variations during anaesthesia induction.

E Weiss1, E Gayat2, V Dumans-Nizard1, M Le Guen1, M Fischler3.   

Abstract

BACKGROUND: Standard non-invasive arterial pressure (AP) measurements are discontinuous. By providing non-invasive beat-to-beat AP measurements, Nexfin™ might limit duration of intraoperative hypotension and hypertension. We assessed the ability of Nexfin™ to detect AP variations by comparing its trending ability with invasive AP monitoring.
METHODS: Thirty-one subjects undergoing elective surgery under general anaesthesia were included. During induction, simultaneous pairs of AP measurements were collected every 5 s from the Nexfin™ finger sensor and a homolateral radial artery catheter. Magnitude and time lags of AP variations from baseline to nadir and peak were calculated for both methods. Concordance analysis was performed by the Bland-Altman method (for comparison of repeated measures when appropriate).
RESULTS: Nexfin™ detected 100% of AP changes with the median delays of 0 s (-13 to 7) and 0 s (-5 to 12) for nadir and peak, respectively. Bias [limits of agreement (LOA)] of systolic AP (SAP) variations was -0.5 mm Hg (-31.2 to 30.2) and -9.4 mm Hg (-31.3 to 12.6) from baseline to nadir and from baseline to peak, respectively. For 3479 analysed paired measurements, bias was -3.8 and -8.8 mm Hg for SAP and diastolic AP, with LOA of (-36.0 to 28.5) and (-29.8 to 12.3), respectively.
CONCLUSIONS: Nexfin™ detects AP variations accurately and can be a useful warning device during anaesthesia. However, it is not interchangeable with invasive monitoring, given the large LOA between the two measurements. CLINICAL TRIAL REGISTRATION: NCT01658631.
© The Author [2014]. Published by Oxford University Press on behalf of the British Journal of Anaesthesia. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Keywords:  anaesthesiology; arterial pressure monitors; monitoring, intraoperative

Mesh:

Year:  2014        PMID: 24771806     DOI: 10.1093/bja/aeu055

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  5 in total

1.  Non-invasive arterial pressure monitoring revisited.

Authors:  Frederic Michard; Daniel I Sessler; Bernd Saugel
Journal:  Intensive Care Med       Date:  2018-03-07       Impact factor: 17.440

2.  Anesthetic management for cesarean section in a patient receiving transplacental treatment of fetal tachyarrhythmia: a case report.

Authors:  Nobuhiro Tanaka; Tomoaki Fujii; Niina Mikami; Yuka Uchinami; Hitoshi Saito; Yuji Morimoto
Journal:  JA Clin Rep       Date:  2019-05-07

3.  Perioperative hemodynamic optimization using the photoplethysmography in colorectal surgery (the PANEX3 trial): study protocol for a randomized controlled trial.

Authors:  Marc-Olivier Fischer; Anne-Lise Fiant; Mariam Boutros; Frédéric Flais; Tzetan Filipov; Stéphane Debroczi; Léa Pasqualini; Toufiq Rhanem; Jean-Louis Gérard; Lydia Guittet; Jean-Luc Hanouz; Arnaud Alves; Jean-Jacques Parienti
Journal:  Trials       Date:  2016-03-22       Impact factor: 2.279

4.  Validation of noninvasive continuous arterial pressure measurement by ClearSight System™ during induction of anesthesia for cardiovascular surgery.

Authors:  Tadashi Tanioku; Akari Yoshida; Yuichi Aratani; Keisuke Fujii; Tomoyuki Kawamata
Journal:  BMC Anesthesiol       Date:  2020-07-20       Impact factor: 2.217

5.  Noninvasive continuous monitoring versus intermittent oscillometric measurements for the detection of hypotension during digestive endoscopy.

Authors:  Anh-Dao Phan; Arthur Neuschwander; Guillaume Perrod; Gabriel Rahmi; Christophe Cellier; Bernard Cholley
Journal:  PLoS One       Date:  2020-10-02       Impact factor: 3.240