Xiangjun Qiu1, Shuang-li Zheng2, Yingfei Wang1, Rong Wang1, Lei Ye3.
Abstract
In this study, a simple, rapid and sensitive ultra performance liquid chromatography-tandem mass spectrometry method is described for the determination of glipizide in human plasma samples using carbamazepine as the internal standard (IS) from bioequivalence assays. Sample preparation was accomplished through protein precipitation with methanol, and chromatographic separation was performed on an Acquity BEH C18 column (2.1 mm × 50 mm, 1.7 μm) with gradient profile at a flow rate of 0.4 mL/min. Mass spectrometric analysis was performed using an QTrap5500 mass spectrometer coupled with an electrospray ionization source in the positive ion mode. The multiple reaction monitoring transitions of m/z 446.1 → 321.0 and m/z 237.1 → 194.2 were used to quantify for glipizide and IS. The linearity of this method was found to be within the concentration range of 10-1,500 ng/mL for glipizide in human plasma. Only 1.0 min was needed for an analytical run. The method was applied to a bioequivalence study of two drug products containing glipizide in human plasma samples.Entities:
Mesh:
Substances:
Year: 2014 PMID: 24771054 DOI: 10.1093/chromsci/bmu023
Source DB: PubMed Journal: J Chromatogr Sci ISSN: 0021-9665 Impact factor: 1.618