Jinghui Zhu1, Mingyou Qin1, Shaopu Liu1, Zhongfang Liu1, Jidong Yang2, Xiaoli Hu3.
Abstract
A flow injection analysis (FIA) system combined with dual-wavelength overlapping resonance Rayleigh scattering (DWO-RRS) has been established and validated for rapid determination of malachite green (MG) and its metabolite in fish samples. Under experimental condition, MG would react with Erythrosin (Ery) to form ion-association complexes, resulting in the occurrence of two RRS peaks and a dramatic enhancement of RRS intensity. The maximum RRS peaks were located at 286 nm and 337 nm. It is noted that the increments of both of these two peaks were proportional to the concentration of MG. The detection limit of DWO-RRS was 1.5 ng/mL, which was comparable to several reported methods. Moreover, the results of real sample analysis exhibited an acceptable recovery between 97.5% and 103.6%, indicating that the method had good reproducibility.Entities:
Keywords: Dual-wavelength overlapping; Erythrosin; Flow injection analysis; Leucomalachite green; Malachite green; Resonance Rayleigh scattering
Mesh:
Substances:
Year: 2014 PMID: 24769380 DOI: 10.1016/j.saa.2014.03.111
Source DB: PubMed Journal: Spectrochim Acta A Mol Biomol Spectrosc ISSN: 1386-1425 Impact factor: 4.098