Effect of phaco tip diameter on efficiency and chatter.

Aabid M Farukhi1, Brian C Stagg1, Cecinio Ronquillo1, William R Barlow1, Jeff H Pettey1, Randall J Olson2, Isha Gupta, Jason D Jensen.   

Abstract

PURPOSE: To evaluate 3 phacoemulsification tips of different sizes and determine which size is most efficient in lens fragment removal using 3 ultrasound (US) approaches.
SETTING: John A. Moran Eye Center Laboratories, University of Utah, Salt Lake City, Utah, USA.
DESIGN: Experimental study.
METHODS: Porcine lens nuclei were formalin-soaked for 2 hours then divided into 2.0 mm cubes; 1.1 mm, 0.9 mm, and 0.7 mm phaco tips were used with torsional and micropulsed US. The 1.1 mm tips were unavailable for torsional US, so 0.9 mm and 0.7 mm tips were used. Efficiency (amount of time for lens removal) and chatter (number of lens-fragment repulsions from the tip) were determined.
RESULTS: The mean phacoemulsification efficiency was highest with the 0.9 mm tip for all US variations. There were statistically significant differences between the 0.9 mm and 0.7 mm tips with micropulsed US (0.8 seconds ± 0.29 [SD] versus 1.4 ± 0.93 seconds; P=.0112) and transversal US (0.8 ± 0.17 seconds versus 1.4 ± 0.89 seconds; P=.0065). There was no significant difference between 0.9 mm and 0.7 mm tips with torsional US or between the 1.1 mm and 0.9 mm tips with micropulsed or transversal US; however, trends were identical, with 0.9 mm tips performing better than 0.7 mm and 1.1 mm tips.
CONCLUSION: With all 3 systems, the 0.9 mm tip was most efficient, with the fewest outliers and smallest standard deviation. FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.
Copyright © 2014 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2014        PMID: 24767913     DOI: 10.1016/j.jcrs.2013.09.021

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  4 in total

1.  The effect of increasing power when grooving using phacoemulsification.

Authors:  Rhett S Thomson; Brian A Bird; Lance A Stutz; Joshua B Heczko; Ashlie A Bernhisel; William R Barlow; Brian Zaugg; Randall J Olson; Jeff H Pettey
Journal:  Clin Ophthalmol       Date:  2019-04-12

2.  Optimization of the Oertli CataRhex 3® phacoemulsification machine.

Authors:  Lance A Stutz; Joshua B Heczko; Brian A Bird; Rhett S Thomson; Ashlie A Bernhisel; William R Barlow; Brian Zaugg; Randall J Olson; Jeff H Pettey
Journal:  Clin Ophthalmol       Date:  2019-04-16

3.  Optimizing Tip Diameter in Phacoemulsification of Varying Lens Sizes: An in vitro Study.

Authors:  Aniket Ramshekar; Joshua Heczko; Ashlie Bernhisel; William Barlow; Brian Zaugg; Randall Olson; Jeff Pettey
Journal:  Clin Ophthalmol       Date:  2021-11-17

4.  Optimization of Phacoemulsification Tip Gauge on the Oertli CataRhex3 in an in vitro Setting.

Authors:  John S Peterson; Michael R Christensen; Caitlynn Cooper; Christina Mamalis; Ashlie A Bernhisel; Brian Zaugg; William R Barlow; Emilie Ungricht; Randall J Olson; Jeff H Pettey
Journal:  Clin Ophthalmol       Date:  2022-04-08