Influence of the intravenous contrast media on treatment planning dose calculations of lower esophageal and rectal cancers.

Jabbari Nasrollah, Molazadeh Mikaeil, Esnaashari Omid, Seyed Siahi Mojtaba, Zeinali Ahad1.   

Abstract

BACKGROUND: The impact of intravenous (IV) contrast media (CM) on radiation dose calculations must be taken into account in treatment planning.
PURPOSE: The aim of this study is to evaluate the effect of an intravenous contrast media on dose calculations in three-dimensional conformal radiation therapy (3D-CRT) for lower esophageal and rectal cancers.
MATERIALS AND METHODS: Seventeen patients with lower esophageal tumors and 12 patients with rectal cancers were analyzed. At the outset, all patients were planned for 3D-CRT based on the computed tomography (CT) scans with IV contrast media. Subsequently, all the plans were copied and replaced on the scans without intravenous CM. The radiation doses calculated from the two sets of CTs were compared.
RESULTS: The dose differences between the planning image set using intravenous contrast and the image set without contrast showed an average increase in Monitor Units (MUs) in the lower esophageal region that was 1.28 and 0.75% for 6 and 15 MV photon beams, respectively.
CONCLUSION: There was no statistical significant difference in the rectal region between the two sets of scans in the 3D-CRT plans. The results showed that the dose differences between the plans for the CT scans with and without CM were small and clinically tolerable. However, the differences in the lower esophageal region were significant in the statistical analysis.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24762502     DOI: 10.4103/0973-1482.131465

Source DB:  PubMed          Journal:  J Cancer Res Ther        ISSN: 1998-4138            Impact factor:   1.805


  6 in total

1.  Influence of the contrast agents on treatment planning dose calculations of prostate and rectal cancers.

Authors:  Sahel Heydarheydari; Negin Farshchian; Abbas Haghparast
Journal:  Rep Pract Oncol Radiother       Date:  2016-05-06

2.  Dosimetric study of Hounsfield number correction effect in areas influenced by contrast product in lungs case.

Authors:  Yassine Oulhouq; Dikra Bakari; Deae-Eddine Krim; Mustapha Zerfaoui; Abdeslem Rrhioua; Soufiane Berhili; Loubna Mezouar
Journal:  Rep Pract Oncol Radiother       Date:  2021-08-12

3.  Evaluation of raw-data-based and calculated electron density for contrast media with a dual-energy CT technique.

Authors:  Daisuke Kawahara; Shuichi Ozawa; Kazushi Yokomachi; Toru Higaki; Takehiro Shiinoki; Yoshimi Ohno; Yuji Murakami; Kazuo Awai; Yasushi Nagata
Journal:  Rep Pract Oncol Radiother       Date:  2019-08-20

4.  Energy Dependence of Measured CT Numbers on Substituted Materials Used for CT Number Calibration of Radiotherapy Treatment Planning Systems.

Authors:  Reza Mahmoudi; Nasrollah Jabbari; Mehdi Aghdasi; Hamid Reza Khalkhali
Journal:  PLoS One       Date:  2016-07-08       Impact factor: 3.240

5.  Intravenous contrast-enhanced cone beam computed tomography (IVCBCT) of intrahepatic tumors and vessels.

Authors:  Cynthia L Eccles; Regina V Tse; Maria A Hawkins; Mark T Lee; Douglas J Moseley; Laura A Dawson
Journal:  Adv Radiat Oncol       Date:  2016-01-26

6.  Intravenous contrast media in radiation therapy planning computed tomography scans - Current practice in Ireland.

Authors:  Shane Minogue; Charles Gillham; Maeve Kearney; Laura Mullaney
Journal:  Tech Innov Patient Support Radiat Oncol       Date:  2019-12-16