Fatemeh Nabavizadeh1, Navin C Nanda2, Amitoj Singh2, Carmen Mateescu3.
Abstract
Six female patients aged from 19 to 73 years, with ostium secundum atrial septal defect underwent closure procedure with Amplatzer septal occluder device. Three-dimensional Echocardiography (3D-TEE) was done during the procedure or one day after the procedure. 3D-TEE provides incremental value over Two- dimensional trans-esophageal echocardiography in measuring Amplatzer septal occluder disc sizes and correlates well with manufacture device size. 3D-TEE will surely prove to increase the technical efficiency and it will become an important tool for the interventionists for periprocedural evaluation of device closures.Entities:
Keywords: Amplatzer Septal Occluder; Atrial Septal Defect
Year: 2012 PMID: 24757602 PMCID: PMC3987407
Source DB: PubMed Journal: Int Cardiovasc Res J ISSN: 2251-9130
Figure 1.Size of LA Side of Device by QLAB.
Comparison between the Results Were and the Manufacturer’s Left and Right Atrial Discs, and Waist Size.
| Age, Sex | Type Of ASD And Device | Observer | ASD Measurements (Mm) | 3 D Tee Measurements (Mm) From Device | ASD Device Size (By Manufacturer) | 2D TEE Bubble Study | Residual Shunt On 3D TEE | Rim Size | |
|---|---|---|---|---|---|---|---|---|---|
| (BY 3D TEE) | LA | RA | WAIST/LA/RA (mm) | ||||||
| 64,F | Aneurysmal Fenestrated | A | NA | 25.2 | 24.6 | ||||
| CribriformASO (25mm) | B | 25.6 | 25.4 | 25 mm CribriformASO | POSITIVE | Positive One DayAfter Procedure | |||
| N/A/25/25 | |||||||||
| 53,F | Secundum ASD | A | 17.4 | 34 | 30.2 | 20 mm ASO | Not Done | Negative | ASD-TV=13.4 mm |
| ASO (20 mm) | B | 19.9 | 34.8 | 30.8 | 4/34/30 | ||||
| 73,F | Secundum ASD | A | 17.3 | 32.4 | 28.8 | 18 mm ASO | Not Done | Negative | ASD-AV=4.7 mm |
| ASO (18 mm) | B | 16.6 | 31.8 | 28.4 | 4/32/28 | ASD-SVC=8.7 mm | |||
| ASD-IVC=7.7mm | |||||||||
| 40,F | Secundum ASD | A | 34.3 | 52.8 | 45.8 | 36 mm ASO | Not Done | Negative | ASD-AV=4.1 mm |
| ASO(36 mm) | B | 31.2 | 52.6 | 46.8 | 4/52/46 | ASD-SVC=7.8 mm | |||
| 30.08 mm-BSD | ASD-IVC=8.6 mm | ||||||||
| 42,F | Secundum ASD | A | 23.5 | 42.8 | 39.2 | 28mm ASO | Not Done | Negative | ASD-AV=4 mm |
| ASO(28 mm) | B | 28.2 | 42.6 | 38.6 | 4/42/38 | ||||
| 19,F | Secundum ASD | A | 34 | 54.6 | 47 | 38 mm ASO | Not Done | Negative | ASD-AV=3.2 mm |
| ASO(38 mm) | B | 32.6 | 54.4 | 48.4 | 4/54/48 | ASD-SVC=8.5 mm | |||
aA and B are two different readers to minimize inter observer error measurements.
bLA;Left Atrium; RA:Right Atrium
*In this patient due to fenestrated septum we did not measure ASD size before the procedure (Cribriform ASO device does not have a waist and has two equal discs). with cribriform ASO device both left and right atrial sides of the disc were equal and disc size was according to LA or RA side, whereas with other ASO devices LA side was larger than RA side and disc size was proportionate to the waist. In this patient the residual shunt disappeared on follow up.
#Malposition of ASD occluder was detected(over the tricuspid valve).Emergency removal of ASD occluder and patch repair of ASD with Bovine pericardium was done.
Figure 2.Size of Ra Side of Device by QLAB.